WARNINGS
This symbol on the instrument indicates that the user
should refer to the manual for operating instructions.
•
Shipping boxes containing an Analyzer may
weigh over 40 lb —necessitating a two-person lift
technique when handling. Such boxes are labeled
with this warning, supporting OSHA standards
intended to protect employees from injury.
•
If this instrument is used in a manner not specified
by Veolia, the safety protection provided by the
instrument may be impaired.
•
To avoid injury and to ensure accurate analysis
results, ensure that the Analyzer door (On-Line
TOC Analyzer) is closed and latched or side
panels (Laboratory and Portable TOC Analyzers)
are on and in place during normal operation.
•
The Analyzer requires an A/C mains supply of
100-240 volts and requires the correct power cord
for safe operation. Use only the supplied power
cord with the Analyzer, which meets this A/C
mains supply requirement.
•
For safety reasons and to avoid the possibility
of corrupting the database, ALWAYS follow this
procedure when shutting down the Analyzer. First
stop analysis AND stop the operating system
(using the Shut down button in the Analyzer's
firmware) BEFORE pressing the Analyzer's power
switch to Off or unplugging the power cord,
•
Any operation requiring access to the inside of
the Analyzer, including installation of maintenance
items, could result in injury. To avoid potentially
dangerous electrical shock, take the following
precautions. BEFORE opening the Analyzer, stop
any analysis, shut down the operating system
(using the Shut down button in the Analyzer's
firmware) and turn off power and disconnect from
the power supply.
•
Electrical conduit and output and alarm
connections should be installed by a qualified
electrician. ESD protection should also be
exercised.
•
This is a Class A product. In a domestic
environment, this product may cause
electromagnetic interference. Take adequate
measures to correct the interference.
•
Hazardous reagents (ammonium persulfate and
phosphoric acid) are used in the Analyzer. Before
installing a reagent, read the corresponding
Material Safety Data Sheet (MSDS) for proper
handling precautions and spill or leak procedures.
The MSDS are contained in the pouch located on
the outside of the reagent shipping box.
© 2023 Veolia. All rights reserved.
32
•
Installation of reagents and most Analyzer
maintenance requires access to the inside of the
Analyzer. To avoid potentially dangerous shock
do this. BEFORE opening the Analyzer, stop any
analysis and shut down the operating system
(using the Shut Down button in the Analyzer's
firmware), then turn off power and disconnect
from the power supply.
•
To avoid exposure to the chemical reagents, wear
acid-resistant gloves and safety goggles.
•
Reagent cartridges are for single-use only. Do
NOT refill. Any refilling or reusing of reagent
cartridges will void all Analyzer and parts
warranties and nullify any performance claims.
•
After removing the plug on the BLOW MOLD
type reagent cartridge, keep the cartridge in an
upright position before attaching and detaching
the reagent supply line to keep chemical from
dripping out and avoid exposure to the chemical
reagent. Save the plug after removing to later
reinsert into the nozzle inlet after removing the
cartridge from the Analyzer during a cartridge
replacement process.
•
To protect against accidental exposure to ultra-
violet radiation, do not operate the UV lamp
outside of its protective housing.
•
(On-Line and Portable TOC Analyzers) Water in
the iOS System may be hot. Before inserting a
vial into the iOS System after operating in On-Line
mode, slide the door open and wait 30 seconds
to allow sample to completely drain. Inserting a
vial before draining can result in hot water spray
projecting upward out of the iOS System.
•
(Laboratory and Portable TOC Analyzers) For
continued protection against fire hazard, replace
fuse with the same type and rating.
(On-Line TOC Analyzer) This symbol indicates
the protective earth terminal (ground) for the
Analyzer.
The iOS System and vial ports contain sharp
needles designed to pierce the septa of sample
vials. Do NOT put fingers or inappropriate
materials into the iOS System or vial port.