Optima Prone / English
3.3.1. For issues that can be checked and solved by user, such as low pressure alarm in head, body
and foot areas, there'll be corresponding image. (see the table below)
(1). Low Pressure Alarm
Alarm will be indicated when the pump detects low pressure condition within the mattress.
NOTE: Under the alternating mode, for patient support, the system will first inflate and hold
the cells with the right pressure level, keep monitoring the cells with low pressure condition.
The system will remain in this state until the low pressure condition in the air cells is solved.
NOTE: The pump can seal the air inside the mattress, which temporaily keeps the patients
from bottoming out during low pressure condition. Patient should be moved if the power
failure issue cannot be solved soon.
(2). Low battery volume
NOTE: When low battery volume image is shown on the display, plug the power cord into the
wall outlet immediately.
CAUTION: Pump system will stop running but display the alarm image and buzzer when low
battery volume <10% is detected. To restore the system to the operating mode, plug the
power cord into the wall outlet, and press the power key to restart the pump.
Code No.
01, 02, 05
03, 04, 06
07
19
Diagram
Description
Low pressure alarm: body section
Low pressure alarm: foot section
Low pressure alarm: head section
Position sensor is not connected
9