Guidance and manufacturer's Declaration – Electromagnetic immunity
The Flocare
Infinity
enteral feeding pump is intended for use in the electromagnetic environment specified below. The customer
®
or the user of the Flocare
Immunity test
Power frequency (50/60
Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Recommended separation distances between portable and mobile RF communications equipment and the
The Flocare
®
Infinity
enteral feeding pump is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Flocare
minimum distance between portable and mobile RF communications equipment (transmitters) and the Flocare
mended below, according to the maximum output power of the communications equipment.
Rated Maximum output
power of transmitter
(W)
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in Watts (W) according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
403259675_r0.indd 32
403259675_r0.indd 32
Infinity
enteral feeding pump should assure that it is used in such an environment.
®
IEC 60601 test level
Compliance level
30 A/m
30 A/m
Flocare
Infinity
®
®
Separation distance according to frequency of transmitter (m)
150 KHz to 80 MHz
d = 1.2
CD
P
0.12
0.38
1.2
3.8
12
Electromagnetic environment – guidance
Power frequency magnetic fields should be at
levels characteristic of a typical location in the typi-
cal commercial or hospital environment.
enteral feeding pump
Infinity
can help prevent electromagnetic interference by maintaining a
80 MHz to 800 MHz
d = 1.2
CD
P
0.12
0.38
1.2
3.8
12
-32-
®
Infinity
as recom-
800 MHz to 6 GHz
d = 2.3
CD
P
0.23
0.73
2.3
7.3
23