4.4 Placement
Place the thermometer between the eyebrows, at a distance
within 5cm from the center of the forehead;
In non-contact mode, the blue pilot light will point to the area
you are aiming at. If the eyebrow area is covered by hair,
sweat or dirt, please clean it in advance to improve the
accuracy of the reading.
Keep the thermometer and forehead still when measuring, as
movement will negatively affect the temperature reading.
4.5 Mode switch
Toggle the slide switch to select the Body mode / Calibration
mode.
Body mode, i.e. operating mode and adjust mode, the
measured result represents the equivalent temperature of
the reference measurement part.
The calibration mode, test mode, represents the temperature
directly measured by the sensor and is used to verify the
accuracy of the laboratory.
4.6 Unit switch
In shutdown mode, press and hold the button "Memory", it
first displays "---- M", which flashes along with the
symbol "M"; press and hold the memory button for more
than 4 seconds, it displays "---- C + human icon + sound
icon"; press and hold the button for 6 seconds until ℃ or °F
appears, press the button "Memory" to free convert
between ℃ to °F. It defaults to the current setting after
rebooting, while it will save previous settings in case of
power failure
4.7 Sound switch settings
In shutdown mode, press and hold the button "Memory", it
first displays "---- M", which flashes along with the
symbol "M"; press and hold the memory button for more
than 4 seconds, it displays "---- C + human icon + sound
icon"; if the sound is on currently, a "bi" will be heard
when entering this mode.
Release the button at this time, it enters the sound setting
mode; press the button "Memory" to change the sound
once; a "bi" will be heard and the sound icon goes out
when the sound is enabled; press the button again to
9
change the sound, the sound icon comes on and there is
no "bi" heard as the sound is disabled.
Note:
The equipment will shut down automatically if there is no
button press for 60 seconds. It defaults to the current
setting after rebooting, while it will save previous settings
in case of power failure.
4.8 Body temperature measurement
Toggle the slide switch to the body mode, press the
button "ON / OFF measurement" to turn on the forehead
thermometer, align the thermometer sensor to the
position between eyebrows within 5cm from the forehead,
press the button "ON/OFF measurement", at this time,
the distance focusing light comes on, quickly adjust the
appropriate distance (preferably when two concentric
circles are overlapping), when a "bi" is heard about 1S
later (no "bi" heard if the sound is disabled), it means
that the body temperature has been measured, with the
result displayed on the LCD screen;
Note:
When the measured temperature is below 37.5℃, the
LCD backlight is green.
When the measured temperature is between 37.5℃-38.5
℃, the LCD backlight is yellow, reminding of low fever.
When the measured temperature is higher than 38.5 ℃,
the LCD backlight is red, reminding of high fever with
sound of "Bi", "Bi", "Bi".
4.9 Object temperature measurement
Toggle the slide switch to the calibration mode, press the
button "ON/OFF measurement" to turn on the thermom-
eter, align the thermometer sensor to the object to be
measured, press the button "ON/OFF measurement", at
this time, the distance focusing light comes on, quickly
adjust the appropriate distance (preferably when the
focusing light spot is within the pea grain size), when a "bi"
is heard about 1S later (no "bi" heard if the sound is
disabled), it means that the surface temperature has
been measured, with the result displayed on the LCD
screen.
10