Electrical Data - Benning MM 5-1 Instrucciones De Servicio

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 41
7.

Electrical data

Note: The measurement accuracy is stated as the sum of
-
a relative proportion of the measurement value and
-
a number of digits (i.e. numerical steps of the last place).
This measurement accuracy applies for a temperature of 18 °C to 28 °C and a
maximum relative humidity of max. 80 %.
The measurement value is obtained as a true RMS value (AC coupling) and
displayed as such. Square-wave signals (> 100 Hz) are not specified. With
non-sinusoidal curves, the value displayed is less accurate. This results in an
additional error for the following crest factors:
Crest factor of 1.0 to 2.0, additional error + 1.0 %.
Crest factor of 2.0 to 2.5, additional error + 2.5 %.
Crest factor of 2.5 to 3.0, additional error + 4.0 % (applies to up to 4000 digits).
7.1 DC voltage ranges
The input resistance is 10 MΩ
Overload protection: 600 V
Measuring range
6.000 V
60.00 V
600.0 V
7.1.1
DC voltage ranges mV DC
The input resistance is 10 MΩ
Overload protection: 600 V
Measuring range
600.0 mV
7.2 AC voltage ranges
The input resistance is 10 MΩ parallel < 100 pF.
Overload protection: 600 V
Measuring
OL indication
range
600.0 mV
6.000 V
60.00 V
600.0 V
*
Measuring range from ≥ 2.0 mV
7.3 DC current ranges (BENNING MM 5-2)
Overload protection:
-
15 A (600 V AC/ DC) fuse, 50 kA, rapid on 10 A input
Max. measuring time:
-
3 minutes at 10 A (pause > 20 minutes)
-
15 secondes at 20 A (pause > 20 minutes)
Measuring range
6 A
10 A
7.3.1
µA DC current ranges (BENNING MM 5-2)
The input resistance is approx. 3 kΩ.
Overload protection: 600 V
Measuring range
400.0 µA
4000 µA
7.4 AC current ranges (BENNING MM 5-2)
Overload protection:
-
15 A (600 V AC/ DC) fuse, 50 kA, rapid on 10 A input
Max. measuring time:
-
3 minutes at 10 A (pause > 20 minutes)
-
15 secondes at 20 A (pause > 20 minutes)
10/ 2015
AC/DC
OL indication
6.600 V
66.00 V
660.0 V
AC/DC
OL indication
660.0 mV
AC/DC
Resolution
660.0 mV
0.1 mV
6.600 V
0.001 V
66.00 V
0.01 V
660.0 V
0.1 V
OL indication
6.600 A
20.00 A
AC/DC
OL indication
440.0 µA
4400 µA
BENNING MM 5-1/ 5-2
Resolution
0.001 V
0.01 V
0.1 V
Resolution
0.1 mV
Measurement accuracy
in frequency range 45 Hz - 500 Hz (sine)
± (2.5 % of reading + 5 digits)*
± (1.0 % of reading + 3 digits)
± (1.0 % of reading + 3 digits)
± (1.0 % of reading + 3 digits)
Resolution
0.001 A
0.01 A
Resolution
0.1 µA
1 µA
Measurement accuracy
± (0.5 % of reading + 3 digits)
± (0.5 % of reading + 3 digits)
± (0.5 % of reading + 3 digits)
Measurement accuracy
± (0.5 % of reading + 5 digits)
Measurement accuracy
± (1.5 % of reading + 3 digits)
± (1.5 % of reading + 3 digits)
Measurement accuracy
± (0.9 % of reading + 5 digits)
± (0.9 % of reading + 5 digits)
23
Tabla de contenido
loading

Este manual también es adecuado para:

Mm 5-2Mm 6-1Mm 6-2

Tabla de contenido