Recommended separation distances between portable and mobile RF communi-
cations equipment and the Penguin Pediatric Nebulizer Compressor.
The Penguin Pediatric Nebulizer Compressor Nebulizer is intended for use in an electro-
magnetic environment in which radiated RF disturbances are controlled. The customer or
the user of the Penguin Pediatric Nebulizer Compressor Nebulizer can help prevent electro-
magnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Penguin Pediatric Nebulizer Compressor
Nebulizer as recommended below, according to the maximum output power of the com-
munications equipment.
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance din metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where Pis the maximum output power rating of the transmit-
ter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propa-
gation is affected by absorption and reflection from structures, objects and people.
Rev.2.08.10.15
14
...........................................................................................................................................................
DRIVE COMPRESSOR/NEBULIZER MODEL MQ6003, MQ6003R, MQ6004, MQ6005, MQ6005R
...........................................................................................................................................................
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2 √ P
0.12
0.38
1.20
3.79
12.00
m
80 MHz to 800 MHz
d = 1.2 √ P
0.12
0.38
1.20
3.79
12.00
800 MHz to 2,5 GHz
d = 2.3 √ P
0.23
0.73
2.30
7.27
23.00