What it means
Caused by colour
bar contamination
or foreign test strip.
The ambient
temperature is less
than 10 °C (50 °F)
The ambient
temperature is
over 40 °C (104 °F)
Test result is lower
than 20 mg/dL
(1.1 mmol/L)
Test result is higher
than 600 mg/dL
(33.3 mmol/L )
Action required
Please insert a new glucose test strip
and perform your test again. If "Sun"
is blinking, avoid direct sunlight and
re-test. Contact your local represen-
tative or the manufacturer, if the
problem persists.
Place the meter at a temperature
between 10-40 °C (50-104 °F) for
more than 30 minutes and test
again.
Place the meter at a temperature
between 10-40 °C (50-104 °F) for 30
minutes or longer, and then test
again.
Check the accuracy of your blood
glucose test strip by performing a
glucose control solution test. If the
control solution test result is in a normal
reading, re-test your blood sample two
or three times. If "Lo" persists, consult
your doctor immediately.
Check the accuracy of your blood
glucose test strip by performing a
glucose control solution test. If the
control test result is in range, re-test
your blood sample two or three times.
If "HI" persists, consult with your
doctor immediately.
Message
What it means
There are no
readings stored
in the meter
Not enough
readings in the
memory to display
designated
average.
Low battery
Battery is dead, or
The glucose test
there is a problem
meter is not
powered ON
with the meter
Glucose test meter
does not start after
being applied
blood sample.
There may be a
Results are
problem with the
inconsistent
glucose test strip
Action required
Battery needs replacing.
Change the battery and if the
problem persists, contact your local
representative or the manufacturer.
Please insert a new test strip and
perform your test again.
Please insert a new test strip and
perform your test again.