en - Areas of application
Cleaning the
A 800 tubular filter
Replacing
filter plates
68
Cleaning or replacing filters
The tubular filter must be cleaned at least once a week.
Follow the cleaning instructions in the operating instructions for the
tubular filter.
Filter plates must be replaced approx. every 2 weeks or after
20 reprocessing cycles have been completed.
Unscrew the holder completely.
Dismantle the holder and dispose of the filter plate .
Rinse out the lower part thoroughly.
Insert a new filter plate.
Damaged or used filter plates must not be inserted under any
circumstances.
Screw the upper and lower parts of the holder back together and fit
the holder back onto the injector manifold.
Do not insert a filter plate into the holder if the injector manifold has
been fitted with an A 800 tubular filter.