12 Operating the unit

12.3 Set the target temperature

From 25–80 °C/77–176 °F.
Press one of the Plus/minus buttons below the temperature display to set the re-
quired target temperature. Press the button repeatedly until the required temperature
flashes on the display. Hold the button down to increase/decrease the duration more
quickly.
IMPORTANT If a temperature limit between 40–60 °C/104–140 °F has been set, see
the Change parameters section for information on changing the temperature limit.
Longer periods of ultrasound operation heat the cleaning fluid in addition to the heat-
ing function. This can lead to the fluid being hotter than the set target temperature.

12.4 Set the ultrasound mode

Degas, eco, sweep, pulse or dyn (dynamic)
Press the Mode button to set the ultrasound mode. Press the button repeatedly until
the required ultrasound mode is shown on the display.

12.5 Set auto-degas

Use this function to degas a fresh batch of cleaning fluid.
One of the ultrasound modes, eco, sweep, pulse or dyn, has been set.
The unit is not currently in ultrasound operation.
Hold down the Start/Stop button until the ultrasound mode degas flashes on the dis-
play.
Press the Start/Stop button again to start the auto-degas function.
Alternatively, save auto-degas as a programme. See the section Save programmes
for more information.
Degas flashes slowly on the display (degassing takes approx. 10 min.). Once auto-de-
gas has been completed, the set ultrasonic cleaning mode will start and be shown on
the display.

12.6 Set Temperature-controlled start

Use this function to start ultrasound operation automatically once the target temperature has been
reached.
IMPORTANT When cleaning sensitive objects, please note that the ultrasound is switched on briefly
multiple times during the heating phase.
The required target temperature has been set.
Ensure that the required target temperature is higher than the current actual temperature. If not, al-
low the fluid to cool.
The unit is not currently in ultrasound operation.
Hold down the Heating button until the Temperature-controlled start symbol flashes
on the display.
20 / 28