Screen displays
Video laryngoscope
Screen
Start screen
Live image
Connection screen
Video error screen
18/124
Meaning
The device switches
on.
The device displays
the video image in real
time.
The remaining opera-
ting time is shown in
minutes next to the
charging indicator in
the top display area.
This animation indi-
cates that the handle
and the display unit
are not connected
with one another.
This screen shows a
video error.
MED 235340
Action
–
–
Connect the handle and the display unit.
Disconnect the handle and the display
unit and reconnect the two components.
The live image is then displayed again.
If this error persists, check the electrical
contacts for soiling and, if necessary, clean
them carefully with a clean and lint-free
cleaning cloth.
If you do not notice any improvement, take
the device out of operation and contact the
manufacturer.