Cleaning And Maintenance - HEBU medical TITANIUM II Instrucciones Para El Uso Y El Servicio

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

EN
Additional remarks:
Because different materials are used, every plaster or plastic cast has different properties.
While working, try to determine the optimum speed for the respective material.
A lower speed is advantages when cutting through plastic casts. This prevents the plastic from
starting to melt due to friction heat generated during cutting.
Where possible, avoid cutting through plaster casts without wadding when using the HEBU
oscillating
Injuries can result if the skin has become stuck to the plaster, and so does not oscillate with
the
Where a plaster cast close to the skin is preferred, we recommend laying a knitted or felt strip
along one side of the arm or leg and drawing a line precisely above this strip on the surface of
the
When sawing open the cast, this line can be used as guiding mark.
When applying a plaster cast following surgery, the position of the operation wound should be
marked on the surface of the plaster cast. This will allow a window to be cut subsequently in
the plaster for removing stitches and observation of the wound.
Cutting a window in the plaster also simplifies the extraction of bone nails and also facilitates
any necessary punctures or incisions.
Metal rails inserted in the plaster cast should not be cut.
While serving, check the retaining screw for a firm fit and if necessary tighten.
If the motor is overloaded, the device is automatically switched off. The fuse (see fig. 9 on
page 28, item 110) is located on the electronic circuit board (see fig. 7 on page 28,
item 110) and can be quickly and easily exchanged after unscrewing the housing.
Used and blunt saw blades place excessive stress on the motor and should consequently be
continuously checked and exchanged in good time.
To ensure even wear of the saw blade, open the retaining screw and turn the saw blade
by 90°.

7 Cleaning and maintenance

Danger of injury through unintentional
activation. Pull out the plug before any cleaning
The following measures should be performed once a week
or more often if the saw is used frequently:
Clean the cooling slots (see fig. 5) to ensure sufficient
ventilation for cooling the motor.
Clean the motor compartment from the outside with a jet of dry compressed air.
Wipe disinfection is suitable for motor housing, cable, power plug and all accessories, such as saw
blades and an open-end wrench. For this purpose, we recommend products with a pH-value within 9-10,
e.g. MediClean forte from Dr. Weigert.
It is important to ensure that the device is disconnected from the power supply and the disinfecting liquid
does not enter the interior of the motor.
24
cast
plaster
cast
or main-tenance work.
cutting
saw
using
Oscillo-Saw TITANIUM II
saw.
blade.
indelible
ink.
Fig. 5
Tabla de contenido
loading

Este manual también es adecuado para:

Hb 8884-aHb 8884-bHb 8884-gHb 8884-sHb 8884-v

Tabla de contenido