P4
Selects the range within which the scale may be zeroed. Note that the
Zero Range
indicator must be in standstill to zero the scale
P5
Sets the level at which motion is detected by comparing the present
Motion Band
display update with the previous one. If motion is not detected for two
seconds or more, the scale is in standstill and can process a Print or Zero
command. Maximum value varies depending on the local regulations
P6
Averages weight readings to produce higher stability. The higher the filter
Digital Filter
setting the greater the stability but the slower the indicator's response
time. Choose SLOW unless a very fast response is needed
P7
Selects the desired formula which determines the point at which the
Overload
indicator shows overload. All selections are based on the primary unit
Limit
selected in P8
P8
Selects the primary base unit to be used in the calibration process. Also
Default Unit
the default unit for normal operation.
"1" = primary unit is lb
P9
Determines the desired weight increments. Value should be conformed to
Display
legal requirements.
Divisions
P10
Determines the location of the decimal point
Decimal Pt
P11
Places the indicator into the zero calibration routine. Scrolling down with
Zero
the [ZERO] key one level begins the procedure
Calibration
P12
Places the indicator into the span calibration routine. Scrolling down with
Span
the [ZERO] key one level begins the procedure
Calibration
P13
Actuates the function that allows the user to view both the zero and span
View
calibration value. The values displayed in this function are valid only after
Calibration
calibration (P11 and P12) have been successfully completed. Scrolling
down with the [ZERO] key one level begins the procedure
P14
Allows the user to key in a known zero calibration value in case of memory
Key-in Zero
loss in the field. Scrolling down with the [ZERO] key one level begins the
procedure
P15
This sub menu will reset all parameters in the "P" and "S" menu to the
Factory
default setting. USE WITH CAUTION!
Reset
PS-IN202 (SS) Indicator
All manuals and user guides at all-guides.com
"FS" = Full Scale in primary units
"2" = primary unit is kg
- 9 -
100%√
1.9%
0.25d
1d√
3d
5d
10d
FAST
SLOW√
FS
FS+ 2%√
FS + 1d
FS + 9d
1√
2
1√
2
5
0√
0.0
0.00
0.000
0.0000
00
Press [ZERO] key to
begin sequence
Press [ZERO] key to
begin sequence
Press [ZERO] key to
begin sequence
Press [ZERO] key to
begin sequence
Press [ZERO] key twice
to begin sequence