POWERLIFT 150/135 Mini/175 User manual
6.7. Transport and storage
All necessary precautions must be taken to ensure the safe transport of the hoist
and its accessories, avoiding knocks and dust.
During transport, the hoist must be in the low position, with its functions discon-
nected and the brakes on the wheels on. Additionally, it must be secured with
straps and protected against water and humidity (75%), at a temperature between
-20ºC and +50ºC.
6.8. Troubleshooting
PROBLEM
The actuator is
not working
The battery will
not charge
The actuator
stops
Noisy hinges
The patient cannot be
lowered
The mast has too
much clearance with
the mast base
The wheels do not
rotate or brake properly
50
CAUSE
The red emergency stop is
activated
The battery is connected to the
electricity mains for recharging
The battery is low
The battery is not properly
connected
The hand control or actuator cables
are not properly connected
Hand control and/or actuator failure
Mains failure
Malfunctioning battery and/or
charging base
Hoist overloaded
The battery is low
Lubrication is required
Actuator and/or hand control
failure while the boom was raised
The grey plastic pieces of the mast
base are been removed
Dirt
REMEDY
Turn the red button clockwise to
deactivate
Unplug the black power supply
cable from the mains
Recharge the battery
Remove the battery and put it
back in place
Insert the cables correctly to
the bottom
Replace the hand control and/or
the actuator
Check mains supply
Replace the battery and/or the
charging base
Reduce the load and wait a few
minutes
Recharge the battery
Apply lubricant to the hinges
Lower slowly the patient turning
the collar on the actuator
(see 4.3. above)
Put the grey plastic pieces on the
mast base again.
Clean the wheels