Technical Data - Bresser 7004040 Manual De Instrucciones

Ocultar thumbs Ver también para 7004040:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

TECHNICAL DATA

Power supply
Built-in battery
Record mode
Record storage
Response Time
Carbon dioxide
measurement range
Accuracy
Resolution
Temperature measuring
range
Temperature resolution
Temperature accuracy
Humidity measuring range
Humidity resolution
Humidity accuracy
Work condition
10
AC/DC power adapter
DC 5V 2000 mA
3.7V lithium battery
• Manual measurement
• Immediate measurement
12700
Within 1 seconds of any
change in CO
2
0 - 9999 PPM
± 70 PPM ±3% readings
1 PPM
-10° to +70°C (14° to 158°F)
0.1°C
±0.3°C
0 to 99.9% RH
0.1%
±3%
0 - 50° C (0 - 85% RH)
Storage condition
Dimensions
Weight
STAND OR WALL MOUNT
You can either mount the device on the wall (2) or use the
supplied stand (6). To use the stand, you must first remove
the screw (5) on the back of the unit and then use this screw
to attach the stand to the back. Tighten the screw hand-tight.
CHARGING
The device has a built-in rechargeable lithium battery.
When the low battery symbol [
screen, please charge the unit using the AC power adapter.
Connect the micro USB cable to the unit, then to the AC
power adapter and then to the wall socket. Alternatively,
connect the unit to a computer USB port for charging.
CALIBRATION
°C/°F
a. The general environment is at 400ppm
b. Make sure the device is shut down
°C/°F
ALM
°C/°F
c. Press the
buttons and power button at the same time
ALM
d. Release the power button and then release the
°C/°F
ALM
REC
buttons - > it will show 400ppm on screen
ALM
REC
MODE
REC
MODE
REC
MODE
°C/°F
-20 - 60° C (0 - 90% RH)
ALM
107x107x33 mm
210 g
REC
MODE
] is displayed on the
Tabla de contenido
loading

Tabla de contenido