CORRECT PIPETTING with the ErgoOne
The volume is set at the factory to the nominal volume for the ErgoOne
changed quickly and easily. See page 8.
Quick start in the standard pipetting mode
1.
Attach the tip
2.
Aspirate liquid
3.
Discharge liquid
4.
Eject tip
10
· ErgoOne E · User Manual
®
Use the correct tip according to the volume range or the color code. Ensure
that the tip is securely seated. When using the flexible pipette shaft, attach an
alternative ejector adjustment clip if necessary.
Pipette tips are disposables items!
Hold the pipette vertically and immerse the tip
2 to 3 mm into the liquid.
Press the pipetting key once
to aspirate the liquid into the
tip. The arrow in the display
points upwards to indicate
the aspiration of liquid.
NOTE:
To avoid the intake of air, leave the tip immersed
into the liquid for approx. 1 sec.
After the liquid has been aspirated, the arrow in the
display points downwards to indicate discharging.
Hold the pipette at an angle between 30º and 45º,
place the tip against the vessel wall.
Press the pipetting key again
and the liquid is completely
discharged including au-
tomatic blow-out. Wipe
pipette tip against the vessel
wall.
Hold the pipette shaft over
a suitable disposal container
and the press the tip ejec-
tion key.
NOTE:
ISO 8655 prescribes rinsing the pipette tip once with the sample liquid prior
to the actual pipetting process.