B-CURE SPORT Instrucciones De Uso página 8

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 43
8. Control panel and troubleshooting
Display
Status
Batteries charging
Batteries fully charged
The icon-
is displayed to indicate the device is currently operating
Full battery power
Low battery power, two beeps will sound after startup as a reminder to recharge. Charge the batteries
No battery power remains. Charge the batteries immediately
The device will shut down automatically after 10 beeps due to lack of battery power. Charge the batteries
Batteries are not charging. Switch the device off then on again to fix this error.
If the problem persists, switch off the device, remove the batteries and then reinsert them.
If the problem persists, ship or deliver personally the device to the service center.
The indication "LOW" is displayed when laser power level is below the required level.
Treatment should be stopped. Cause may be one of the following:
1. Battery power is too low.
2. Diode lifespan has reached its end and should be changed at the service center.
3. The device is operated in a different environment than for which it was designed, e.g.
temperatures higher than 40°C. Charge the batteries, switch on the device and see if the
writing "LOW" has disappeared. If not, verify that room temperature is not exceptionally high, and if
necessary move away from the hot area. If not higher than 40°C and the writing "LOW" is still displayed -
ship the device to the service center for diode change.
Additional signals for errors of this type:
Hardware/software error has occurred. Switch the device off, then on again, to fix this error.
If the problem persists, switch off the device, remove the batteries and then reinsert them.
If the problem persists, ship or deliver personally the device to the service center.
8
Tabla de contenido
loading

Productos relacionados para B-CURE SPORT

Tabla de contenido