Calibration
Your instrument was carefully tested and calibrated before being shipped from the factory. For greatest ac-
curacy, we recommend factory re-calibration every 6-12 months. Call customer service at (630) 543-3747. If
you wish to do calibration yourself, follow these procedures.
1.
To activate calibration mode, turn the unit on and press and hold both the On/Off button and the
Adjust-Down button (making sure to press the Adjust-Down button first so the unit will not turn off).
The led will blink in amber rapidly for about five seconds, and turn solid green. Release the On/Off
button and the Adjust-Down button, the led will then blink amber at which point only the pen being
adjusted will show on the display.
2.
To raise the unit of measurement, press the Adjust-Up button. To lower the unit of measurement,
press the Adjust-Down button. Pen home switches between the red and blue pens on a two pen unit
and
stores the current adjustment value.
3.
When calibration is complete, press the On/Off button. Calibration is stored in memory even after you
turn the unit off. User calibration information will not be lost if AC power fails.
4.
Note: After two hours, if no buttons are pressed, the unit will time out of user calibration and resume
normal operation. The user calibration is stored separately from the factory calibration. If you wish to
cancel out your user calibration, simply enter calibration mode and toggle through the steps without
adjusting displayed readings. Exit by pressing the On/Off button. You have now restored factory
calibration.
Note: It is recommended that you use a controlled chamber when calibrating temperature and humidity.
Calibrating in an open room is not recommended as humidity can vary greatly within a very small area.
Rev. 03/08