Pre-Installation Requirements - KROHNE SMARTPAT COND 1200 Manual

Tabla de contenido
SMARTPAT COND 1200

3.3 Pre-installation requirements

PRECAUCIÓN!
¡
Do not drop the device! Handle the device with care!
Never touch or scratch the electrodes of the sensor.
Store the sensor in its original packaging in a dry, dust-free location. Keep it away from dirt. If
necessary, clean it. See cleaning procedure en la página 28
Do not make any mechanical modifications to the sensor (electrodes shortened, drilled, bent
or scratched). This can result in the loss of proper functionality, as well as the rights under
the device warranty.
The sensor must be suitable for the temperature, pressure and medium conditions which are
specified (including chemical resistance).
The device must not be heated by radiated heat (e.g. exposure to the sun) to a electronics
housing surface temperature above the maximum permissible ambient temperature. If it is
necessary to prevent damage from heat sources, a heat protection (e.g. sun shade) has to be
installed.
INFORMACIÓN!
¡
Está disponible un software DTM específico para el sensor para el uso con FDT PACTware
software DTM es gratuito y disponible en CD (suministrado) o puede descargarse en la página
web de KROHNE (Descargas).
The required steps are explained in the following sections.
Figura 3-1: Desembalaje del sensor
Unpacking the sensor
• Remove by gently twisting and pulling the protective cap from the sensor 1.
• Lay the sensor on a soft ESD mat or soft paper tissue 2.
• Leave the protection cap of the VP connector, as long as the sensor is not connected to the
cable.
05/2017 - 4006081701 - MA SMARTPAT COND1200 R01 es