A separate window will open automatically:
6. Now manually enter the actual size in mm of the specified distance in the Actual dimension field.
In the image above, a distance of 1 mm was measured and thus "1" entered in the field.
The software will now automatically calculate the selected magnification.
7. Confirm your entry with OK. The calculated magnification will then be displayed in the upper
right-hand corner of the window (both in the Image Preview and the main program window) and
defined as a value for all subsequent photos taken.
• Following calibration, reliable measurements may be taken on all subsequent photos as
i
long as the focus setting is not changed.
• If you adjust the focus for a photo using the focus knob, you will need to perform a
calibration again.
• Should you ever need to, you can also manually specify the magnification for an individual
photo or future photos.
The software offers different options:
Reset Magnification
Select Reset Picture Magnification to restore the selected magnification value of an individual
photo back to zero.
The magnification value displayed in the upper right-hand corner of the Image Preview will
be restored to None for the open photo.