Rs 485 Communication Description; Transmission Channel Characteristics; Message Structure - Varian Turbo-V70 Manual De Instrucciones

Ocultar thumbs Ver también para Turbo-V70:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 19

RS 485 Communication Description

Communication RS 485 serial port connections

Transmission channel characteristics

levels:
RS485
baud rate:
9600/4800/2400/1200/600
programmable
character lenght:
8 bits
parity:
none
stop bit:
1 bit
protocoll:
master (PC) / slave (converter)
max. devices:
32

Message structure

(request and answer have the same format)
The master system (PC) starts every session sending
the following message to the slave units connected:
<STX> + <ADDR> + <WINDOW> +
<COMMAND> + <DATA> + <ETX> +
<CRC>
where :
<STX>=
0x02
<ADDR> =
0x80 (for RS 232 and
RS 422 only)
<ADDR> =
0x80 + device number (0...31)
0xFF: brodcasting command
(recognized by all the devices, it
doesn't implicate any answer)
(for RS 485 only)
<WINDOWS>=
' 000 '...' 999' window number
the meaning of the window
depends to the device type
<COMMAND>=
0x30: window value reading
0x31: window writing
<DATA> =
alphanumeric ASCII string
containing, in the case of writing
operation,the parameter to input
into the window addressed by
the field <WINDOW>This field
may have variable length
according to the data type
contained in the window where
you are working in. In the case of
reading request of a window, the
data field doesn't exist.
<ETX>=
0x03
<CRC>=
XOR among all the characters
following <STX>=(with exception
of <STX>), including the end
character <ETX> hexadecimally
encoded by two ASCII characters.
When a slave device is addressed by the master :
1)
In case of reading request of the value contained in
a window, the slave answers a string equal to the
one sent by the master but in addition there is the
field
<DATA>
window.The format of the field <DATA> depends to
the window type.
The different types are :
Lenght
logic
1
analog
6
alphanumeric
max 10
EXAMPLES:
Command : START
Source
: PC
Destination : Inverter
02
80
30
STX
ADDR
WINDOW
Source
: Inverter
Destination : PC
02
80
06
STX
ADDR
ACK
59
TECHNICAL INFORMATION
containing
the
value
Characters permitted
'0'=OFF
,'1'=ON
'-', '-','0'...'9'
( to the right with '0')
' '...'_'
03
30
30
31
31
WR
ON
ETX
03
38
35
ETX
CRC
87-900-883-01 (C)
of
the
42
33
CRC
Tabla de contenido
loading

Este manual también es adecuado para:

969-9509

Tabla de contenido