Troubleshooting Information; Emc Information - HERSILL V7g Instrucciones De Uso Y Mantenimento

Equipo de aspiracion medica
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

D. Disinfection / Sterilization:
Disinfected/Sterilized jars should be dated and stored in clean and dry
areas to avoid recontamination
Avoid to put in cntact the parts with the autoclave chamber, because
the material can be damaged. It is recommended to use a basket.
All parts can be processed pursuant one of the next 2 methods:
Steam Sterilization in gravity autoclave:
Polycarbonate (PC) jars: (max. 40 times) at 121ºC and 1.3k kg/cm
Polysulfone (PSU) jars: (max.100 times) at 132-136ºC and 2.4 kg/cm
High level Disinfection:
With Glutaraldehyde, Cidex: 2% solution at ambient temperature during
60 minutes. Rinse thoroughly with sterile water, by immersing it
completely in three separate copious volumes of water. Each rinse
should be a minimum of one minute. Do not reuse the water. Finally,
allow to dry.
E. Inspection , reassembly and perform the function test pursuant Chapter 5.

8. Troubleshooting information

FAULT
The V7g unit
does not
operate
The V7g unit
does not work
from its battery
The V7g unit
operates, but
with little flow
or little vacuum

9. EMC Information

V7g meets the IEC 60601-1-2 EMC requirements for medical devices.
Medical electrical equipment needs special precautions regarding EMC and needs
to be installed and put into service according to the EMC information provided in the
accompanying documents. Portable and mobile RF communications equipment can
affect medical electrical equipment.
All manuals and user guides at all-guides.com
ENGLISH
during 25 minutes, drying time: 30 minutes.
during 15 minutes, drying time: 30 minutes.
CONDITION
Control Unit not turned on
Wrong operating voltage
Mains not connected
External power indicator lit
Defective Fuse
Battery status LEDs is not
lit
Floater valve is closed
Collection jar full
Defective connection from
the Unit to the jar
Patient Suction Tubing
twisted or blocked
Turn Main Switch „ON/OFF"
Check mains voltage output
Connect mains to the Unit
Send the unit to repair
Check fuse and replace it
Place the V7g unit on charge.
If does not operate after
charging, replace the battery
Unplug the Filter/Vacuum
tubing to release the vacuum
Remove and replace the jar
Install the Vacuum Connector
Tubing correctly
Untwist, clear or replace the
tubing
10
2
ACTION
2
Tabla de contenido
loading

Tabla de contenido