Troubleshooting

PROBLEM
LCD display
shows abnormal
result
12
Display messages in LCD
PROBLEM POSSIBLE CAUSE SOLUTION
Low battery
symbol
Er 0
Er 1
Er 2
Er 3
Er 4
Er 5
Er 6
Er 7
Er 8
Er 9
Er A
No response
when you press
button or load
batteries.
All manuals and user guides at all-guides.com
POSSIBLE CAUSE
The cuff position is not correct
or cuff is not properly tightened
Body posture was not correct
during testing
Talking, arm or body movement,
irritation, excitement or tension
during testing.
Irregular heartbeat (arrhythmia)
Low battery
Pressure system is unstable
before measurement
Fail to detect systolic pressure
Fail to detect diastolic pressure
Pneumatic system blocked or cuff
is too tight during inflation
Pneumatic system leakage or cuff
is too loose during inflation
Cuff pressure above 300mmHg
More than 3 minutes with cuff
pressure above 15 mmHg
EEPROM accessing error
Device parameter checking error
MCU self-verify error
Pressure sensor parameter error
Incorrect operation or strong
electromagnetic interference.
SOLUTION
Apply the cuff correctly
and try again
Review the "BODY POSTURE
DURING MEASUREMENT"
sections of the instructions
and re-test.
Re-test when calm and without
speaking or moving during
the test
It is inappropriate for
people with serious arr-
hythmia to use this blood
pressure monitor.
Change batteries
Don't move and try again.
Apply the cuff correctly
and try again
Wait 5 minutes and
measure again. If the monitor
is still abnormal, please
contact the local distributor
or the manufatcurer.
Take out batteries and
and reinstall after 5 minutes.