POWER-ON SELF TEST
The illumipro-10 automatically performs an internal Power-On Self Test (POST) each time the
instrument is powered-on. The POST Test confirms that software and hardware components of the
system are performing as expected. Successful completion of the POST Test will be followed by an
audible tone. POST Test failure will be indicated by an Error Code. Additional information regarding
Error Codes is located in Appendix III of this manual.
OPTICS SYSTEM VERIFICATION
Calibration of the illumipro-10 is not required. Verification of the OPTICS SYSTEM for each Block must
be performed monthly to ensure proper function. Verification is completed using the Red Verification
Standard included with the illumipro-10.
Stepwise instructions for Optics System Verification are
provided on the illumipro-10 display, utilizing the SERVICE MODE menu.
illumipro-10 Operator's Manual: ENGLISH