INHALATION THROUGH MOUTH
1. Accessories: connectors and connection tube (as shown in drawing C/a)
- Install them to the main unit
- Insert the mouthpiece (figure C/b) and put the other side into your month. Start the treatment..
- In addition it is possible to use the appliance in a more simple way, as shown in image C/c.
In order to change the nebulisation rate, just press the power switch.
2. If you should fill uncomfortable during treatment, please stop immediately.
PLEASE INHALE THE MIST SLOWLY BY YOUR MOUTH
INHALATION THROUGH NOSE
Please install the mask
Kindly note that your nose should be inside the mask, as shown by image D/a
1. Please press the power switch in order to adjust the nebulisation rate.
2. Once the nebulisation has ended, unplug the main unit.
Note: After every treatment switch off the appliance for 3 minutes.
Attention. The aerosol is designed for spraying medicinal solutions with a diameter of particles equal to that
indicated in the section "TECHNICAL CHARACTERISTICS". Medicinal solutions with a superior diameter are not
sprayed effectively.
CLEANING
Before cleaning the appliance, always plug out.
After each use always clean accurately all the accessories to remove every possible residues of medicinal. Wash
carefully each used component under running warm water. If necessary add a little amount of not aggressive
cleansing solution. Rinse
abundantly, drain any water and dry thoroughly into a clean ambient. Use a soft, damp cloth to wipe the body, the
cover and the tank clean. Drain any water and dry
thoroughly into a clean ambient. Don't use soap or powder that can damage the surface. To avoid infections or
chemical reactions, don't use twice the same drug holder medicine cup.
NEVERWASH THE BODY UNDER RUNNING WATER.
How to disinfect the product
Disinfect every day the product how it is explained underneath.
Wash used accessories with chemical disinfectant, easily searchable at a drugstore and suitable to plastic material.
To correctly dose the disinfectant read the instruction shown on the disinfectant itself. Rinse carefully under running
water and dry.
Maintenance activities other then cleaning and parts disinfection can be carried out only by staff authorized.
13