Indicator / Symbol Description
Alarm
ALARM text and alarm number with red back-light will
display if alarm is triggered. An audio alert will also
sound for 20 minutes before automatically shutting off.
Empty (outline)
A full charge has three (3) solid bars. Two (2) bars =
battery indicator
50%. One (1) bar = 25%. Outline only = <13%
CaL
Displayed in the ID field, this indicates a temperature
display issue.
HI
Cryogenic chamber temperature is above 49°C; no data
is recorded
HL
The connection point between thermocouple wires and
the instrument electronics module ("cold junction")
is above its operating limit of 49 °C; "Not valid"
temperature will be reported in the log file
Key
Displayed in the three (3) digits ID Code area; this
indicates one or more keys are lodged in a depressed
position.
LL
The connection point between thermocouple wires and
the instrument electronics module ("cold junction")
is below its operating limit of 49°C; "Not valid"
temperature will be reported in the log file.
LO
Chamber temperature is below -199°C; "Not valid"
temperature will be reported in the log file.
Low Memory
Available memory is running low, and will only
record up to 6 additional hours of log data when first
displayed.
not SET
"not SET" indicates that the date/time are not set; if
(Date/Time)
time is not programmed, "not SET" is displayed at all
times.
OFF
OFF indicates that the internal basket is above -20°C.
If OFF is displayed while unit is on, unit will turn OFF
automatically within one (1) minute if no button activity
is detected.
OP
Service issue: Thermocouple detachment detected.
Service
SERVICE message indicates a technical support issue
such as thermocouple detachment (OP) or corrupted
temperature correlation (TMP), stuck key (KEY),
EPROM reading error (EP), thermocouple chip error
(TIC). Message will automatically turn OFF if issue is
resolved.
TMP
Displayed in the three (3) digit ID Code area, this
indicates the pre-loaded data needed for temperature
indication is corrupted.
All manuals and user guides at all-guides.com
Resolution
Resolve alarm as
needed.
Replace batteries as
needed.
Contact Technical
Support.
Add LN2 as needed.
Contact Technical
Support.
Depress keys to
dislodge.
Contact Technical
Support.
Contact Technical
Support.
Download log data
immediately to free
up space.
Program date and/or
time in Set-Up.
Fill with LN2 as
needed.
Contact Technical
Support.
Contact Technical
Support.
Contact Technical
Support.
I
English
5
Tabla de contenido
loading

Tabla de contenido