EN
English
Connection cable
Pressure resis-
tance
Measurement con-
ditions
Storing conditions
Characteristics
when delivered
Measuring
ranges,
resolution
7
Length
Diameter
Smallest allowed
bend radius
Plug type
Sensor with connection
cable
The IDP 761-C meets the requirements according to article 3(3) of the
directive, 97/23/EC ("pressure equipment directive").
Conductivity measuring
range
Temperature range
Max. admissible overpres-
sure
Minimum depth of immer-
sion
Maximum depth of immer-
sion
(at temperature)
Operating position
Recommended storing
method
Storage temperature
Temperature responding
behavior
Accuracy of the temperature
sensor
Measuring ranges, resolution, accuracy
Measured parameter
ӕ [µS/cm]
ӕ [mS/cm]
1.5 m
4.3 mm
Fixed installation: 20 mm
Flexible use: 60 mm
Socket, 4 pins
5
IP 68 (2.5 x 10
Pa or 2.5 bar)
1 S/cm ... 500 mS/cm
-5 ... 70 °C (100 °C) 23 ... 158 °F (212 °F)
IDP 761-C: 2.5 x 10
36 mm
Whole sensor + cable up to 70 °C (158 °F)
Sensor shaft only (=120 mm) up to 100 °C
(212 °F)
Any
In air
0 ... 50 °C (32 ... 122 °F)
t
(99 % of the final value display after)
99
< 20 s
± 0.2 K
Measuring range
0.0 ... 199.9
200 ... 1999
2.00 ... 19.99
20.0 ... 199.9
200 ... 500
IDP 761-C
5
Pa (2.5 bar)
Resolution
0.1
1
0.01
0.1
1
ba77220e01