5. INSTALLATION

5.1 SETTING UP THE LED LIGHT SOURCE
Place the LED light source on a stable surface (cart, counter, stand, etc.).
NOTE
Avoid places where the light source may be splashed with liquid.
Absolutely DO NOT use in any environment with explosive or
flammable gases.
DO NOT block the venting grids of the light source or impede
exhaust air flow. A minimum of 6"/15cm on each side and
12"/30cm at the back is required, 24"/60cm is preferred. The
lightsource should not be placed where its exhaust will influence
other devices, nor where the heated air from other devices will
influence the lightsource.
Make sure the power switch is in OFF position.
Connect AC power cord to the power inlet located on the rear panel of the light
source.
Use only cords provided with the light source.
CAUTION
Plug the AC power cord into a wall outlet using the three-prong plug supplied
with the unit.
To prevent electric shock, connect power cords of peripheral
equipment through medical isolation transformers.
CAUTION
NOTE
When using medical isolation transformer, be sure to check the
transformer power ratings.
Make sure that the power cord is connected to the main power with
three-prong plug (USA use UL2601-1 rated isolation transformers
and/or power strips only).
When used in clinical or residential areas near radio or TV units,
this equipment may be subjected to radio interference.
CAUTION
To avoid adverse electromagnetic effects,
DO NOT operate this equipment near RF energy equipment.
5.2 CONNECTING THE FIBER OPTIC CABLE
Turn intensity control to the minimum setting. Connect the light cable to the instrument
and then plug the light cable endfitting into the port on the front panel.
LIT168 INTERTEST
(English)
Rev.-
Page 9 of 60