To replace the filter, proceed as follows:
1.
Unscrew the luer fitting from one side of the filter and
detach the tubing from the other side.
Note: If re-fitting the existing filter, make sure that filter
orientation is maintained (yellow label on the filter,
is facing the instrument).
2.
Fit a new In-line Hydrophobic Filter (Part No. 66484)
with the yellow label on the filter, facing the instrument.
3.
Attach the luer fitting, with sample line attached to the
filter.
4.
Attach the sample line to the filter.