Regulatory Information
† The ISM (industrial, scientific and medical) bands between 150 kHz and 80
MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz
to 27. 283 MHz; and 40.660 MHz to 40.700 MHz.
‡ Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast, and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the CPRmeter 2 is used
exceeds the applicable RF compliance level above, the CPRmeter 2 should be
observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the
CPRmeter 2.
Recommended separation distances between portable and
mobile RF communications equipment and the CPRmeter 2
The CPRmeter 2 is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of
the CPRmeter 2 can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications
equipment (transmitters) and the CPRmeter 2 as recommended below,
according to the maximum output power of the communications equipment.