10. It is necessary to keep quiet during the measurement to get an
accurate reading. Measurement should be conducted in a quiet
environment at room temperature. Don't eat or smoke before a
measurement.
This device is supplied with the standard cuff which fits arm sizes
8.7-13.8 in. / 22-35 cm.
WARNING: Ensure the cuff size is appropriate for the person
whose blood pressure is being taken.
WARNING: Do not use any cuff other than the original cuff
contained in this kit!
11. Repeated measurements with an interval of 3 minutes are
recommended, so you can calculate the average to get a more
accurate measurement. Atherosclerosis patients may require a
longer interval (10-15 minutes) as elasticity of the patient's vessels
decreases significantly with the disease. 10-15 minute interval is
also applicable for patients suffering from longtime diabetes.
CLASSICATION
ME EQUIPMENT is not intended for use in an oxygen rich
environment or in the presence of flammable mixers.
Internally powered equipment.
Type BF applied part, cuff recognized as applied part.