1 OVERVIEW

Abrasives used for blast cleaning surfaces can be contaminated with
soluble salts due to the source or the re-use of the blasting media.
This contamination can be transferred to the blast cleaned surface and
result in accelerated corrosion conditions and also cause premature
coating failure, if this contamination is not removed prior to applying
the coating.
Testing abrasives for soluble salt contamination can be carried out
quickly and easily using the Elcometer 138 Abrasive Soluble Salt Test
Kit, according to the ASTM D4940 method.
Elcometer 138 Abrasive Soluble Salt Test Kit provides all the
equipment needed to carry out the test in the field or in the laboratory.

2 BOX CONTENTS

Elcometer 138E Conductivity Meter
Standard 1413 µS/cm (1.413 mS/cm) Calibration Solution,
1 x 20ml (0.74 fl oz) Single Use Pouch
Bottle of Pure Water; 1000ml (33.8 fl oz)
Glass Beaker; 100ml (3.4 fl oz)
Glass Beaker; 600ml (20.3 fl oz)
Plastic Measuring Beaker; 500ml (16.9 lf oz)
Funnel
Stirring Rod
Filter Papers; Box of 100
LR44 Alkaline Batteries; x4
Transit Case
User Guide
a
The Elcometer 138E Conductivity Meter included in the test kit measures the conductivity of aqueous
solutions. The meter is NOT designed to measure solids, organic solvents, surfactant, oil, adhesive,
alcohol, strong acids (pH: 0 to 2) or strong alkalis (pH: 12 to 14). The life of the sensor will be extremely
short if these substances are measured.