Periodical Maintenance Checks - Gima 28220 Manual De Instrucciones

Ocultar thumbs Ver también para 28220:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 21
9
Silicone tubes: the number of cycles of sterilization and/or cleaning is strictly linked to the employment of the
said tube. Therefore, after each cleaning cycle, it is up to the final user to verify whether the tube is suitable
for reuse. The component must be replaced if there are visible signs of decay of the material constituting the
said component.
Conical fitting: the number of cycles of sterilization and the number of cleaning cycles is strictly linked to
the employment of the said component. Therefore, after each cleaning cycle, it is up to the final user to verify
whether the fitting is suitable for reuse. The component must be replaced if there are visible signs of decay in
the material constituting the said component.

PERIODICAL MAINTENANCE CHECKS

The TOBI MANUALE SUCTION ASPIRATOR does not need maintenance or lubrication. It is necessary to
check functioning and instrument before every use. Unpack the instrument and always check integrity of
plastic parts, they might have been damaged during previous use. Close the aspirator outlet with your finger.
Press the foot and check vacuum functionally.
Type of fault
1. No aspiration
2. No aspiration
3. The float doesn't close
4. The float doesn't close
5. Low suction
6. No aspiration due
to flow leakage of mucus
7. The Vacuum power
on the patient side is
either very low or absent
Faults 1 - 2 - 3 - 4 - 5 -
6 - 7
If the overfill security system is activated, don't proceede with the liquid aspiration. If the overfill security sys-
tem doesn't work there are two cases:
1
case - If the overfill security system doesn't work the aspiration will be stopped by the bacteriological filter
st
who avoids the liquid penetration inside the device.
2
case - If both the security system and the bacteriological filtrer do not work, there is the possibility that
nd
liquid has leaked inside the device, in this case return the device to GIMA technical service.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING,
PLEASE CONTACT GIMA TECHNICAL SERVICE. GIMA DOES NOT GIVE GUARANTEE IF INSTRUMENT,
AFTER THE TECHNICAL SERVICE CHECKING, APPEARS TO BE TAMPERED.
Cause
Jar Cap badly screwed down
Lid seal not in its seat
The float it's covered by dirty ma-
terial
If the cap has been washed, en-
sure that the float is not partially
detached
Foam inside the jar
Filter blocked
• Protection filter blocked or dam-
aged
• Connection
tubes
kinked or disconnected
• Shut-off valve blocked or dam-
aged
• Pump motor damaged
None
of
the
remedies
achieved the desired results
ENGLISH
Remedy
Unscrewed the cap, then rescrew it cor-
rectly
Unscrew the cap and insert the seal prop-
erly in its seat
Unscrewed the cap, leave the and put in
on autoclave
Fit the float into it's place
Fill the jar to 1/3 full of ordinary water
Replace filter
• Replace the filter
• Replace or reconnect the tubes, check
blocked,
the jar connections
• Empty the jar, or disconnect the tube
from the jar and unblock the shut-off
valve. The unit twill only work in the up-
right position
• Refer to authorised service personnel
has
Contact the seller or GIMA After-sales As-
sistance Service
Tabla de contenido
loading

Tabla de contenido