Applications
The Elite PCTS tester can be used for the following:
• Agriculture
• Aquaculture
• Aquariums and fish farms
• Boiler blow-down
• Car Washes
Warranty
This instrument is supplied with a warranty against manufacturing defects for a period of one year from the date of
purchase.
Return of Items
Authorization must be obtained from your distributor before returning items for any reason. When applying for
authorization, please include information regarding the reason the item(s) are to be returned.
We reserve the right to make improvements in design, construction and appearance of products without notice.
Prices are subject to change without notice.
Self Diagnostic Messages
Batteries are weak and need replacement soon.
stable error
Appears when calibration is attempted but the reading is not yet stable. Wait for the reading
to stabilize or manually confirm the calibration by pressing enter.
buffer error
The buffer is outside of the calibration range.
slope error
The 2
over range
The reading is above the measuring range of tester.
under range
The reading is below the measuring range of tester.
8
• Drinking water Hydroponics
• Ecology
• Electroplating rinse tanks
• Food Sectors
• Labs
or 3
calibration point is not within 80% to 120% slope range.
nd
rd
Insert Sensor Module
Oring ID22XCS3
Oring ID21XCS2
Sensor
Small Tab
Turn anti-clockwise to release
• Printing industry
• Swimming pools
• Verification of reverse osmosis system operation
• Water and wastewater treatment
• Others
Elite PCTS Tester Operating Instructions
Ring