Itm Fluke T5-600 Manual De Instrucciones página 2

Auto Off
45 min
LO
Do not
overtighten
2 AA
+
screw
+
OFF
Warning
Disconnect test leads from power source
before opening battery door.
Débranchez les cordons d'essai avant
d'ouvrir le couvercle de la pile.
Prima di aprire lo sportello del vano della pila,
scollegare i cavetti dal circuito di
alimentazione.
Die Prüfleiter vor dem Öffnen des
Batteriefachs von der Stromquelle trennen.
Desconecte los conductores de prueba de la
fuente de alimentación antes de abrir la
cubierta de la batería.
Antes de abrir a tampa do compartimento das
pilhas, desconecte os terminais de teste da
fonte de alimentação.
Maak de meetkabels los van de stroombron
voordat u de batterijklep opent.
Koble prøveledningene fra strømkilden før
batteridekselet åpnes.
Søgeledningerne skal tages evt. emne, inden
batteridækslet åbnes.
Irrota koestusjohtimet virtalähteestä ennen
paristolokeron avaamista.
Koppla bort mätsladdarna från strömkällan
innan du öppnar batteriluckan.
1.800.561.8187
www.
Specifications
Specifications subject to change without notice.
Calibration
1-year calibration cycle
Temperature
Operating:
-10 °C to +50 °C (14 °F to 122 °F)
Storage:
-30 °C to +60 °C (-22 °F to +140 °F)
Altitude
Operating:
2000 m (6562 ft)
Storage:
10,000 m (32808 ft)
Relative
0 % to 95 %, 5 ° C to 30 ° C
Humidity
(41 ° F to 86 ° F); 0 % to 75 %,
30 ° C to 40 ° C
(86 ° F to 104 ° F); 0 % to 45 %,
40 ° C to 50 ° C
(104 ° F to 122 ° F)
Battery Type
AA (2); 360 hours continuous with
and Life
alkaline; 125 hours continuous with
zinc chloride
RF Field
0.5 % full scale + (specified accuracy)
Specification
at 3 V / m
Enclosure
IP 52 per IEC 60529, no vacuum
Rating
applied
Temperature
0.1 x (specified accuracy) / ° C for <18
coefficient
° C or >28 ° C (< 64.4 ° F or > 82.4 ° F)
± ( [ % of reading] + [number of least
Accuracy
significant digits] )
Accuracy is specified for 1 year after
calibration, at 18 ° C to 28 ° C
(64 ° F to 82 ° F) with relative humidity
to 90 %. AC conversions are ac-
coupled, average responding, and
calibrated to the rms value of a sine
wave input.
Safety ......................... IEC 61010-1: Pollution Degree 2
Electromagnetic
Environment ................ IEC 61326-1: Portable
.com
T5-600
T5-1000
Range
Range
Reso-
Function
Accuracy
(600 V
(1000 V
lution
CAT III)
CAT III)
600 V
1000 V
1 V
± (1.5 % + 2 digits)
rms
45 Hz - 66 Hz
rms
600 V
1000 V
1 V
± (1 % + 1 digit)
100.0 A
100.0 A
0.1 A
± (3 % + 3 digits)
45 Hz - 66 Hz
1000 Ω
1000 Ω
1 Ω
± (1 % + 2 digits)
Replacement Parts
Test Lead Assembly.
Replace only with Fluke double-insulated
PN 648029
leads ( ) See "Accessories" for available
probes.
B
t a
e t
y r
D
o
r o
PN 2091779
B
t a
e t
y r
D
o
r o
S
c
e r
w
PN 8758161
Use only specified replacement parts.
Accessories
H
5 Belt Holster
TP1 Probe Set, Flat-Blade
AC285 SureGrip™ Alligator Clips
AC220 SureGrip™ Alligator Clips
Service Information Sheet, PN 686953
loading

Este manual también es adecuado para:

Fluke t5-1000