Tektronix TDS2MEM Manual Del Usuario página 14

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 97
Using the TDS2MEM Storage Memory and Communications Module
Key Points
Saves All To Files.
fast and easy way to save all of the current oscilloscope information
to files with a single button push. Pushing the PRINT button creates a
new folder and saves the following information in separate files in that
new folder, using the current instrument and file format settings, as
listed in the following table:
Source
CH(x)
MATH
Ref(x)
Screen Image
Settings
.CSV Files.
string that lists the time (relative to the trigger) and voltage values for
each of the 2500 waveform data points. You can import .csv files into
many spreadsheet and math analysis applications.
Screen Image Files.
spreadsheet and word processing applications. The type of image file
you can import depends on you application.
.SET files.
oscilloscope settings.
6
Setting the Print Button to Saves All To Files is a
File name
FnnnnCHx.CSV, where nnnn is an automatically-
generated number, and x is the channel number.
FnnnnMTH.CSV
FnnnnRFx.CSV, where x is the reference memory letter.
FnnnnTEK.???, where ??? is the current Save Image file
format.
FnnnnTEK.SET
CSV (comma-separated value) files contain ASCII text
You can import screen image files into many
SET files contain an ASCII text string listing of the
TDS2MEM User Manual