Picture C. Cup fixing
EQUIPMENT USE:
1. To do viscosity measurements, a stopwatch is needed as well as
other supplied elements.
2. Warming cup up: Start up controller switching on main switch (1).
Select a temperature pushing the SET key (2) and, at same time,
(3) or (4) keys until the display shows required temperature.
3. Filling cup: Block viscosimeter bottom hole with a finger. Fill it with
the liquid as well as it spill by the rim.
4. Expelling surplus liquid: Using level glass drive top liquid paying
attention that any air bubble remain into the cup. Leave glass as
a cover lid. In this moment, bottom hole can be unblocked (liquid
will not fall).
5. It can be leaved some minutes to sure that liquid is enough tem-
pered.
6. Place a vessel under bottom viscosimeter hole. Remove glass
and at same time start up stopwatch. Stop it the first time liquid
trickle cuts.
It is important to remove glass making an horizontal movement
as you can see on picture E.