REFILL DATA
Patient Name ___________________________________________________
Date
Time
Volume mL
1
Volume
Returned
Infused
From Pump
1. Subtract return volume from volume at last refi ll:
(16 mL for CODMAN 3000-16; 30 mL for CODMAN 3000-30; 50 mL for
CODMAN 3000-50).
2. Flow rate per day. Divide volume infused by number of days from last refi ll.
EXAMPLE:
CODMAN 3000-30
Volume
Returned
30 mL Reservoir
From Pump
8. CLOSE STOPCOCK. Disconnect the stopcock
and syringe barrel leaving the needle and refi ll set in
place. Note the returned volume (mL) and record on
refi ll data sticker provided. Discard syringe barrel and
stopcock leaving the needle and refi ll set in place.
9. Expel air from the 10 mL syringe of saline. Attach
the syringe to the proximal end of the refi ll set and
confi rm that the needle is still in contact with the
needle stop. Keep downward pressure on the needle
and inject 5 mL of saline into the pump.
SALINE SOLUTION
10. Release pressure on the plunger and allow the
5 mL injected to return to the syringe. This procedure
reconfi rms correct positioning of the needle. If no fl uid
returns, refer to Troubleshooting Guide for assistance.
Disconnect the syringe from the refi ll set.
CODFR13007xx02svn.indd 3
CODFR13007xx02svn.indd 3
No. of Days
2
Flow Rate
Refi ll substance, concentration and total volume in mL
From Last Refi ll
mL/Day
A
B
Volume
Infused
30 mL – A
2 mL
30 – 2 = 28 mL
11. Expel air from the refi ll syringe. Attach the
appropriate volume syringe to the proximal end of the
refi ll set and confi rm that the needle is still in contact
with the needle stop.
12. Keep downward pressure on the needle and
begin to inject refi ll solution into the pump. Release
pressure on plunger at 5 mL increments and allow
LID
1 mL of solution to return to syringe. This will verify
that the needle is in the correct position and the pump
reservoir is being fi lled. Continue to inject and check
STOPCOCK
needle placement until the syringe is emptied.
CLOSED
PRECAUTION: If no fluid returns to the syringe
upon release of the plunger, DO NOT CONTINUE
TO INJECT REFILL SOLUTION UNTIL YOU HAVE
VERIFIED THE NEEDLE PLACEMENT PER THE
PROCEDURE IN THE TROUBLESHOOTING
GUIDE. Follow steps A, B, C, D per the

Troubleshooting Guide.

13. After injecting the entire contents of the refi ll
syringe, close the clamp on the tubing set and pull the
needle out of the pump septum. Remove drape and
iodine. Apply adhesive bandage to access site.
5 mL
Inform patient of date, time and place for next
Refi ll Volume:
CODMAN 3000-16 ........... 16 mL
CODMAN 3000-30 ........... 30 mL
CODMAN 3000-50 ........... 50 mL
CODMAN
®
3000-16 (16 mL)
®
CODMAN
3000-30 (30 mL)
CODMAN
®
3000-50 (50 mL)
C
D
No. of Days from
Flow Rate
Last Refi ll
B/C
28 days
28 mL/28 days =
1 mL/day
1 mL
1 mL
5 mL
5 mL
5 mL
REFILL SOLUTION
Special Note:
pump refi ll.
Troubleshooting Guide:
If no fl uid returns during pump emptying, or during
the Refi ll Injection Procedure, proceed as follows:
A. Confi rm needle position and that a 22 gauge,
non-coring needle is being used. Check that the
needle is PERPENDICULAR to the pump septum
and is fully depressed and in contact with the
needle stop. If no volume returns proceed with
steps B, C, D.
B. Attach a 10 mL syringe containing saline
solution to the refi ll tubing set.
INCORRECT POSITION
INCORRECT POSITION
NEEDLE STOP
CORRECT NEEDLE POSITION
C. Keep downward pressure on the needle and
inject 5 mL of saline into the pump. Release
pressure on the syringe plunger and allow the
fl uid to return to the syringe. If fl uid returns then
the needle is in the proper position and indicates
that the pump was empty at the start of the refi ll
procedure. Replace the 10 mL syringe with the
syringe containing the pump refi ll solution and
continue with the refi ll procedure (steps 12 and 13).
D. If there is still no fl uid return after injection of
the 5 mL of saline, remove needle and refi ll tubing
set. Flush the refi ll set to confi rm that the system is
patent. Re-insert the needle perpendicular into the
pump septum until it is in contact with the needle
stop. Repeat step C.
E. If there is still no fl uid return after injection of the
5 mL of saline, contact Codman & Shurtleff, Inc. for
24-hour technical assistance at +1 508 880 8156 or
800 660 2660.