The number of autoclave is 20 times in Main China and Japan ONLY due to local
»
approval issue.
Medit will not be responsible for any damage including distortion, blackening etc.
▪
4.2.3
Mirror
The presence of impurities or smudges on the tip mirror may lead to poor scan
quality and an overall poor scanning experience. In such a situation, you should
clean the mirror following the steps below:
▪
Disconnect the scanner tip from the i500 handpiece.
▪
Pour alcohol on a clean cloth or cotton-tipped swab and wipe the mirror. Make sure to
use alcohol that is free of impurities or it may stain the mirror. You can use either ethanol
or propanol (ethyl-/propyl alcohol).
▪
Wipe the mirror dry using a dry, lint-free cloth.
▪
Make sure the mirror is free of dust and fibers. Repeat the cleaning process, if necessary.
4.2.4
Handpiece
After treatment, clean and disinfect all other surfaces of the handpiece except for the
scanner front (optical window) and end (air vent hole).
Cleaning and disinfecting must be done with the device turned off. Use the device only
after it is completely dry.
Recommended cleaning and disinfecting solution:
Denatured alcohol (aka. ethyl alcohol or ethanol) – typically 60-70% Alc/Vol.
The general cleaning and disinfecting procedure is as follows:
▪
Turn off the device using the power button.
▪
Unplug all the cables from the power hub.
▪
Attach the handpiece cover to the front of the scanner.
▪
Pour the disinfectant onto a soft, lint-free and non abrasive cloth.
▪
Wipe the scanner surface with the cloth
▪
Dry the surface with a clean, dry, lint-free and non-abrasive cloth.
CAUTION
▪
Do not clean the handpiece when the device is turned on as the fluid may enter the
scanner and cause malfunction.
▪
Use the device after it is completely dry.
Intraoral Scanner i500
13