Temperature Measurement - Chauvin Arnoux C.A 801 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

3. TEMPERATURE MEASUREMENT

Implementation of the C.A 801
I Connect the K thermocouple to input T1.
I Connect the adapter to the digital multimeter, with the red
terminal connected to the + input and the black terminal
connected to the COM input. Select the V ... or mV ...
calibres.
I Choose the measurement units (°C or °F) using the
selector.
Implementation of the C.A 803
I Connect the K thermocouple(s) to input T1 or/and T2.
I Connect the adapter to the digital multimeter, with the red
terminal connected to the + input and the black terminal
connected to the COM input. Select the V ... or mV ...
calibres.
I Use the selector to choose T1 or T2 or T1-T2 (RCD mode).
I Choose the measurement units (°C or °F) using the
pushbutton.
NB:
Do not use the V elec (Z = 270 kΩ) input on C.A 5210
and C.A 5220 multimeters.
° C
-40°C to 0°C
0°C to 400°C
400°C to 1000°C
°F
-40°F to 32°F
32°F to 400°F
400°F to 1800°F
(1) Of the adapter + K sensor delivered with the instrument (in
the reference range +18°C to +28°C).
Accuracy indicated in relation to the output signal (mV/°C
or °F).
Protection
Accuracy
(1)
(K thermocouple)
+ 0.8 % + 2 mV
60 V ...
+ 0.5 % + 1 mV
+ 0.8 % + 2 mV
Protection
Accuracy
(1)
(K thermocouple)
+ 0.8 % + 4 mV
60 V ...
+ 0.5 % + 2 mV
+ 0.8 % + 4 mV
11
~
24 V
~
24 V
Tabla de contenido
loading

Este manual también es adecuado para:

C.a 803

Tabla de contenido