EN
PALAMIX
Footswitch
Description
The PALAMIX
vacuum pump is a compressed air operated
®
vacuum pump. This generates a vacuum in the mixing
cartridge of the PALAMIX
PALACOS
R+G pro, which allows mixing of bone cement
®
under vacuum.
The PALAMIX
vacuum pump is located in the non-ster-
®
ile area on the floor of the operating room.
Installation and connection
The PALAMIX
vacuum pump is connected with the in-
®
house compressed air supply. The functional range of the
pump is an operating pressure between 5 and 10 bar
(0.5–1.0 MPa) air pressure. In order to allow the use of the
PALAMIX
vacuum pump, it has to be connected to the
®
mixing cartridge of the PALAMIX
or PALACOS
R+G pro. For this purpose, the respective
®
vacuum hose supplied has to be connected via the acti-
vated charcoal filter to the PALAMIX
PALAMIX
vacuum pump can only be used in c onjunction
®
with the PALAMIX
vacuum mixing system or the
®
PALACOS
R+G pro. The respective operating steps are
®
also described here in detail.
vacuum pump
®
vacuum mixing system or the
®
vacuum mixing system
®
vacuum pump. The
®
Maintenance and cleaning
Regular maintenance of the PALAMIX
not required. The pump housing can be cleaned with a
soft cloth without aggressive detergents. Take care here
that no cleaning fluid or water enters the PALAMIX
vacuum pump.
Notes
The PALAMIX
vacuum pump should not be used without
®
a filter element. In the event of malfunctioning of the
PALAMIX
vacuum pump contact your local Heraeus
®
Medical representative.
4
Vacuum hose connection
Compressed air connection
vacuum pump is
®
®