4. Installation and First Use
4.1 Using the appliance for the first time
Remove the parts from the box and check for com-
pleteness. If any parts are missing or damaged,
please contact Customer Service. We recommend
that you keep the packaging for transportation
purposes. Before using the articulator for the first
time, spray with a standard silicone spray. Do
not grease!
4.2 Setting up the appliance (average value) (3.0)
> Loosen knurled thumb screw A and set condyle
path inclination to an average value of 30°.
> Close the centric lock. (4.0)
> Push the vertical guide rod on the vertical guide
rod holder into the zero position (millimetre
scale) and fasten.
> Centre the incisal guide on the vertical guide rod
> Push the long part of the incisal pointer through
the top hole and the short part through the bot-
tom hole on the vertical guide rod as far as they
will go.
4.3 Support rod
The supporting rod is a feature of the ergonomic
design of the articulator. Screw the supporting rod
into the thread in the upper section of the CA 3.0.
When the supporting rod is fitted the upper sec-
tion of the articulator will be supported (opening
stop).
4.4 Mounting in the articulator (1.0 + 2.0)
The following possibilities are available for
mounting in the articulator:
> Direct fixing with plaster thanks to the
split-cast system (1.0)
> Plastic mounting plates, Metal mounting
plates (2.0)
5. Handling, Operation
5.1 Centric position
The CA 3.0 features a precision centric lock which
can be activated in 3 possible positions:
Centric lock (4.0) closed
In this position the CA 3.0 can only be opened and
closed in centric position for occlusion checks to
be performed.
Centric lock (4.0) open
Retrusion lock closed
Protrusive, Bennett, balance movements as well
as immediate side shift (ISS) and all intermediate
movements can be physiologically reproduced in
this position.
Centric lock and retrusion lock open (5.0)
With the lock in this position, the retrusive move-
ment can also be performed.
5.2 Protrusive movement (6.0)
If the centric lock is open, the protrusive move-
ment can be performed as shown in the figure.
The horizontal condyle path inclination can be
adjusted from 0 – 60° by rotating the wheel.
5.3 Lateral and Bennett movement (7.0)
If the centric lock is open, the lateral movements
can be performed as shown in the figure. This
is done by applying backward pressure with the
thumb on one side on the appropriate joint.
5.4 Immediate side shift (ISS) (8.0)
If the centric lock is open, the immediate side shift
can be performed as shown in the figure. This is
done by pressing in the transverse direction on
one side with the index finger.