INTRODUCTION
The model LF-60A is a passive high impedance oscilloscope probe, designed
and calibrated for use on instruments having an input impedance of 1MΩ //
20pF. However, it may be compensated for use with instruments having an input
capacitance of 10pF to 50pF.
COMPENSATION ADJUSTMENT
Connect the probe to the oscilloscope and to a 1kHz square wave source. Switch
the attenuation to the x10 position.
Set the oscilloscope to display two to three cycles and two to six vertical divisions.
Carefully adjust the trimmer to obtain the flat tops to the square wave on the
oscilloscope, according the illustration of Figure 1.
SPECIFICATIONS
Feature
Attenuation
Bandwidth
Rise Time
Input Resistance
Input Capacitance
Compensation Range
Operation Voltage
Safety
Cable Length
INTRODUCCIÓN
El modelo LF-60A es una sonda de osciloscopio pasiva de alta impedancia,
proyectada y calibrada para uso en instrumentos con impedancia de entrada
de 1MΩ // 20pF. Entretanto, puede ser compensado para uso en instrumentos
con capacitancia de entrada de 10pF a 50pF.
AJUSTE DE LA COMPENSACIÓN
Conecte la sonda al osciloscopio y en una fuente de onda cuadrada de 1kHz.
Ponga la llave de atenuación en x10.
Configure el osciloscopio para exhibir de dos a tres ciclos y de dos a seis
divisiones vertical.
Cuidadosamente ajuste el trimmer para obtener una onda cuadrada con bordas
rectas en el osciloscopio, conforme la ilustración de la Figura 1.
LF-60A OSCILLOSCOPE PROBE
Position x10
10MΩ (1MΩ oscilloscope)
Approx. 15pF
10pF to 50pF
600V DC / AC Peak Derating w/ Frequency (Figure 2)
LF-60A SONDA DE OSCILOSCOPIO
10:1
DC ~ 60MHz
5.8ns
Compliance IEC-1010 CAT II
Approx. 1.2m
Position x1
1:1
DC ~ 15MHz
23.3ns
1MΩ (of oscilloscope)
46pF + oscilloscope
capacitance
-