NOTICE
1.Keep the sensor and probe cavity clean before and after
use.
2. To ensure the accuracy of measurement, it is
recommended to start measurement after ten minutes
when carrying the thermometer to a new environment.
3. Wait for 10 minutes to measure body temperature
after measuring extremely high temperature or
extremely low temperature objects.
4. Wait for 5 minutes to start a measurement when a
measuring target (object or human) is from an
environment with enormous difference in temperature.
5. Breeze, water, sweating, cosmetic on forehead may
affect measurement. Do not measure body temperature
in 30 minutes after exercise, bath, or meals.
O
BJECT TEMPERATURE
•
Press measurement button to turn on the
thermometer (fig.9.6).
•
After measured room temperature make sure the
thermometer is under object mode (fig.9.7-9.8).
•
Keep vertical distance at 5cm to 15 cm from object
to measurement probe. Press measurement
button and when it gives a beep measurement is
finished and value is idle in 20 seconds. It will
display "OFF" (fig. 9.9) and gives a "beep" and shut
off automatically.
IACER Srl
9
ENG
MNPG314-00