Calibration
1.
Press TARE/CAL key; the value of the necessary weight will flash on display
Put the calibration weight on the pan. Wait until the value of the calibration weight
2.
stops flashing and is fixed on the display (e.g. 100.00 g, 200.00 g, etc., depending
on the model)
3.
Remove the weight from the pan
The balance will enter weighing mode and will display 0.000g.
4.
Note: Each balance includes the necessary weight to complete external calibra-
tion,but the capacity which is 2kg is not included.
Tare function
1.
Put the receptacle on the weighing pan; its weight will be displayed
2.
Press TARE/CAL key and the reading on the display will automatically be set to
0.00g; tare is complete
3.
If the receptacle is removed from the pan, the corresponding weight, with
negative value will be displayed.
4.
Press the TARE/CAL key again and the value on display will be set to zero again
Weighing unit selection
Press UNIT key to select the desired weighing unit from g, kg, ct, T, TAR, dr, PKT, GN,
TMR, gsm, tlJ, mo, dwt, oz, lb, tlT, ozt, tlH, %
Piece counting function
Press the COU button
1.
Press the TARE/CAL button repeatedly to cycle through different counts. This function
2.
can be used with counts of 10, 20, 50, 100, 150, 200, 250, or 500 pieces.
Put the corresponding number of pieces on the pan, press the PCS button to confirm.
3.
4.
Remove the pieces from the pan and the balance is ready to be used in piece
counting mode.
To return to weighing mode, simply press the COU button again.
5.
RS232 Interface
Balance (9 pins) Printer/ PC (9 pins)
RXD(Input) 2 --------------------------- 3
(Output) 3 ---------------------------- 2
(Ground) 5 ---------------------------- 5
•
The default Baud Rate is 9600 bps
•
The data format is: 10 bits; 0 as start bit, 1 as stop bit, 8 digits
(ASCII code)
•
No odd and even numbers
adjustments Data output format:
1
2
3
4
Type
Space
Space
±
or *
9
10
11
12
Data
Data or
Data or
Data
dot
dot
YPZ Models • All Rights Reserved •
• www.labbox.com
5
6
7
8
Data
Data
Data
Data
13
14
15
16
Unit 1
Unit 2
End
Return
II. Specifications
Technical Parameters:
BYPZ-
Model
100-001
Capacity
0-100 g
Readability
Repeat ability
Linearity
Pan size
Dimensions
280 * 220 *140 mm
Net weight
Output
YPZ Models • All Rights Reserved •
• www.labbox.com
BYPZ-
BYPZ-
200-001
300-001
0-300 g
0-200 g
0.001 g
± 0.002 g
± 0.004 g
Dia. 90mm
2000 g
110/220-240 V