"CAL" icon and calibrating value (75,2% if at 25ºC) are blinking on the LCD with current temperature at the left.
Meter is now calibrating. Wait about 60 minutes until blinking stops, then calibration is completed and it returns
to normal mode.
NOTE: Users can also calibrate either point. To calibrate 33% only, pres CAL/ESC key to exit when 33%
calibration is completed. To calibrate 75% only, press UP or DOWN key within 5 minutes while initializing 33%
calibration. It will skip 33% calibration and enter 75% calibration mode.
Troubleshooting
Problem
Can't power on
Slow response
Error messages
Error text
E01
E02
E03
E04
E07
E11
E17
E31
E34
Specifications
Model
CO
Range
2
CO
Resolution
2
CO
Accuracy
2
CO
Pressure dependance + 1.6% reading per kPa deviation from normal pressure, 100
2
Temperature Range
Temperature Resolution
Temperature Accuracy
Humidity Range
Humidity Resolution
Humidity Accuracy
Warm up
Operating
Storage
Power
Solution
Check if the adaptor is well plugged
Check if the air flow channels on the rear are blocked
Meaning
CO
sensor damaged
2
The value is under range
The value is over range
The original data error results in this error (RH, DP, WB)
Too low voltage to measure CO
Check if the adaptor output is 12V
Retry humidity calibration
Retry CO
calibration
2
Temperature sensor damaged
Humidity sensor damaged
7722
0-2000 ppm
(2001-9999 ppm out of scale)
1 ppm
±50 ppm + 5% rdg (0-2000)
kPa
-10.0 – 60.0 ºC
0.1 ºC/0.1 ºF
± 0.6 ºC/ ± 0.9 ºF
0.0-99.9%
0.1%
±3% (10-90%)
±5% (other range)
30 seconds
0 – 50 ºC, 0 – 95% RH (avoid condensation)
-20 – 60 ºC, 0 – 99% RH (avoid condensation)
12V adaptor
2
9