The contents contained in this manual are subject to change without prior notice. Information furnished by CMI Health is believed to be accurate and reliable. However, no responsibility is assumed by CMI Health for its use, or any infringements of patents or other rights of third parties that may result from its use.
Página 2
User Manual for Pulse Oximeter Instructions for Safe Operation Check the device to make sure that there is no visible damage that may affect user’s safety and measurement performance. It is recommended that the device should be inspected minimally before each use. If there is obvious damage, stop using the device.
Página 3
User Manual for Pulse Oximeter For the individual patients, there should be a more prudent inspecting in the placing process. The sensor can not be placed on the edema and tender tissue. The local law should be followed when disposing of the expired device or its accessories.
Página 4
User Manual for Pulse Oximeter and follows the provisions of the council directive MDD93/42/EEC.
User Manual for Pulse Oximeter Press this key, the default screen can be shifted to display pulse rate (PR) or perfusion index (PI). If on the system setup screen, press it to move the cursor upwards or to the left and adjust parameter values.
User Manual for Pulse Oximeter 1.2 Product Name and Model Name: Pulse Oximeter Model: PC-66L 1.3 Structure It consists of the main unit and an adult SpO probe. 1.4 Features It is lightweight, small in size and easy to carry ...
1)The battery compartment is located at the back of the device and can be open with a coin or a flat-head screwdriver, as shown in Figure 2-1. DO NOT open the battery compartment unless told by CMI tech support. Figure 2-1 2)Insert the metal stand as shown in Figure 2-2.
Página 10
Securely place the new battery in the battery compartment. Do Not attempt unless told by CMI tech support. Safety instruction for operation: Do not throw the device or battery into water, liquid or fire.
User Manual for Pulse Oximeter If the device does not power properly or the battery is damaged, please contact CMI customer service. In order to prolong the lithium battery's using life, please pay attention to the battery maintenance.
Página 12
User Manual for Pulse Oximeter 4. Avoid placing the sensor on the same limb which is wrapped with a cuff for blood pressure measurement or during venous infusion. 5. Do not let anything block the emitting light from the sensor. 6.
User Manual for Pulse Oximeter 4 Operation 4.1 Power on/off the Oximeter Press and hold " " power/back key for 2 seconds to power the device ON or OFF. During measurement, if the "Probe off" status is showing for longer than one minute, and if there is no keypad operation for one minute, then the device will power off automatically.
Página 14
User Manual for Pulse Oximeter Description: " means the device is on mute status, the Sound indicator " user can press recall/mute " " key to turn on the indication sound. During the measurement, over-limit event or probe off event can trigger the alarm. ”: Memory full icon;...
User Manual for Pulse Oximeter 4.3 Display Screen with PI Value On the default display screen, press “ / ” Navigation key to shift screens between default screen and display screen with PI value. The display screen with PI value is shown below. Perfusion index index Perfusion index index Figure 4-2 Display Screen with PI Value...
Página 16
User Manual for Pulse Oximeter Figure 4-3 Setup Menu Screen Screen Description “Setting”: set parameter values, refer to Chapter 4.4.1 for details. “Upload data”: enter into data uploading state, refer to Chapter 4.4.2 for details. "Default": enter into the factory default setting, refer to ...
Página 17
User Manual for Pulse Oximeter Figure 4-4 System Setup Screen Operation Instructions: Patient: two options:"ADU " for adult, and "PED " for pediatric.
Página 18
User Manual for Pulse Oximeter Lo-Limit: SpO low limit setting; range: 50%~99%, the step is 1%. The factory default value for adult is 90% and 95% for pediatric. PR Hi-Limit: High limit setting of pulse rate; range: 100~240bpm. From 100 to 150, the step is 1bpm, and from 150 to 240, the step is 5bpm.
Página 19
User Manual for Pulse Oximeter confirm and exit from date setting. The procedures of adjusting Month value and Day value 4) are the same with Year adjustment. Date Format: yyyy-mm-dd Note: The setting operations of other parameters (such as TIME, PATIENT, RECORDING INTERVAL, POWER SAVING etc.) are the same with date setting.
User Manual for Pulse Oximeter Power saving: power saving setting; two options: "on" and "off". The factory default setting is "on". When the power saving mode is set to "on" option, during the measurement, if there is no key operation for 2 minutes, the screen display will be dim for power saving.
User Manual for Pulse Oximeter Figure 4-5 Connecting Status Screen 4.4.3 Default On the menu screen, select "Default" and press " "key to enter the default setting screen (as shown in Figure 4-6). Press the Navigation " " key to choose "Yes" or "No", and press "...
User Manual for Pulse Oximeter language changes to English. In English version: on main menu screen, select " key to change display "espanol" and then press " language to Spanish. 4.4.5 Version On the menu screen, select “VERSION” and then press “ ”...
Página 23
User Manual for Pulse Oximeter 4.5.1 Data Recall Choose record in the record list, then press “ ”(mode/OK) key, the screen will display trend graph, as shown in Figure 4-9A. Recording interval trend graph PR ordinate (yellow line) P R tren d graph (green line) ordinate Figure 4-9A Trend Graph Display Screen...
Página 24
User Manual for Pulse Oximeter Figure 4-9B Trend Graph Display Screen Figure 4-9C Trend Graph Display Screen 4.5.2 Data Deletion (all device data) On the menu screen, longtime press “ ” key and the records list will appear. At this time, longtime pressing “ ”...
Página 25
User Manual for Pulse Oximeter Figure 4-10 At this time, press "" navigation key to select “Yes” or "No", and press “ ” mode/OK key to confirm or exit. Or short time press " " power/back key to return to record list screen.
User Manual for Pulse Oximeter 5 Technical Specifications A. Display Mode: Color dot-matrix LCD; B. Power Supply: Supply voltage: d.c. 3.6V (Lithium rechargeable battery) Operating current: ≤180mA C. SpO Measurement Transducer: dual-wavelength LED sensor with wavelength: Red light: 663 nm, Infrared light: 890 nm. Maximal average optical output power: ≤...
Página 27
User Manual for Pulse Oximeter Operating Humidity: 30%~80% Atmospheric pressure: 70kPa~106kPa Note: portable and mobile RF communications equipment may affect the performance of the Oximeter. G. Data Update 8 beats moving average for both SpO and pulse rate readings H. Data Storage Recording SpO and pulse rate data every 1/2/4/8 second(s), up to 384–hour records can be stored.
User Manual for Pulse Oximeter The device is not intended for use in the environment with rich oxygen or ignitable gas. Mode of operation: Continuous operation. Electro-Magnetic Compatibility: Group I, Class B 6 Over-limit Indication 6.1 Limit settings low limit setting range: 50% ~ 99%. ...
The alert sound starts (interval is 5 seconds) and lasts for about 1 minute. If the probe is still off, the Oximeter will power off automatically. 7 Packing List 1. The PC-66L oximeter 2. An adult SpO probe 3. A SpO2 extension cord 4.
User Manual for Pulse Oximeter 8 Repair and Maintenance 8.1 Maintenance The expected service life(not a warranty) of this device is 5 years. In order to ensure its long service life, please pay attention to the maintenance; The recommended storage environment of the device: Ambient temperature: -4ºF ~140ºF Relative humidity 10%~95% Atmospheric pressure: 50kPa~107.4kPa...
User Manual for Pulse Oximeter 9 Troubleshooting Trouble Possible Reason Solution Device cord Check the probe to see if the light kept disconnected or source is blinking red light. If not, showing broken circuit. either the connection cord or the “Check sensor is malfunctioning.
User Manual for Pulse Oximeter Improper Please use the suggested sensor sensor/improper according to user’s body type and body type setting usage. Improper sensor or body type setting may cause inaccurate readings. device Please contact customer service. multifunctioning. No signal Device shows If the monitor still shows pulse...
Página 33
There’s a build-in memory storage stored? inside the device that could store up to 384 hours of data. How do I manage CMI Health’s offers a free PC the data? software, Oximeter Data Manager, which could easily download from our website at: www.cmihealth.com/download...
Página 34
User Manual for Pulse Oximeter What computer CMI Health’s Oximeter Data system does the PC Manager software support software support? Windows 7/8/10. How can I transfer CMI Health’s Oximeter Data data from Manager software provides computer to another? with data import and export feature.
User Manual for Pulse Oximeter Appendix Key of Symbols Symbol Description %SpO The oxygen saturation Symbols Perfusion Index on the Pulse rate (Unit: beats per minute) screen Low battery voltage Battery fully charged Speaker mute icon Speaker volume icon Memory full (Pediatric/Adult) Patient type...
Página 36
User Manual for Pulse Oximeter Symbol Description probe connector Power/Back Key Mode/OK Key Recall/Mute Key / Navigation Key Data Interface Symbols CE mark on the Serial number panels Date of manufacture Manufacturer (including address) With Type BF applied part See User Manual Disposal of this device according to WEEE regulations No alarm...
Página 37
User Manual for Pulse Oximeter Pulse Oximetry Knowledge 1 Meaning of SpO is the saturation percentage of oxygen in the blood, so called O concentration in the blood; it is defined by the percentage of oxyhemoglobin (HbO ) in the total hemoglobin of the arterial blood.
Página 38
User Manual for Pulse Oximeter contrast, hemoximeters report fractional oxygen saturation – a percentage of all measured hemoglobin, including dysfunctional hemoglobin, such as carboxyhemoglobin or metahemoglobin. Clinical application of pulse oximeters: SpO is an important physiological parameter to reflect the respiration and ventilation function, so SpO monitoring used in clinical becomes more popularly, such as monitoring the...
Página 39
User Manual for Pulse Oximeter such as nail enamel or color skin care Excessive patient movement Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line Exposure to the chamber with High pressure oxygen ...
Página 40
El contenido de este manual está sujeto a cambios sin aviso previo. La información facilitada por CMI Health se considera que es precisa y fiable. Sin embargo, CMI Health no asume ninguna responsabilidad por su uso, ni por las infracciones en las patentes u otros derechos de terceros...
Página 41
Manual de uso para Pulsioxímetro Instrucciones para su funcionamiento seguro Verificar el dispositivo para comprobar que no haya daños visibles que puedan afectar la seguridad del usuario y a la precisión de la medición. Se recomienda que el dispositivo se examine como mínimo antes de ser usado.
Página 42
Manual de uso para Pulsioxímetro durante el proceso de colocación. El sensor no se puede colocar sobre tejidos en periodo de cicatrización o edemas. Deben respetarse las leyes locales sobre el desguace del dispositivo y sus accesorios una vez finalizada su vida útil. Atención ...
Página 43
Manual de uso para Pulsioxímetro Declaración de Conformidad El fabricante declara que este equipo cumple con las siguientes normas: IEC 60601-1, ISO 9919:2009, ISO 80601-2-61:2011 sigue recomendaciones directiva MDD93/42/EEC.
Página 44
Manual de uso para Pulsioxímetro Indice 1 Vista General................. 1 1.1 Presentación..............1 1.2 Producto Nombre y Modelo..........3 1.3 Estructura................. 3 1.4 Prestaciones..............3 1.5 Uso al que se destina............4 1.6 Entorno de Funcionamiento..........4 2 Instalación de la Batería y el Soporte........5 3 SpO Conexión del Sensor............
Manual de uso para Pulsioximetro 1 Vista General 1.1 Presentación Interfaz de data oporto de cargador Pantalla Teclas de ON/OFF/Atrás Navegación Llamar/Silencio Modo/OK Conector sensor SpO Figura 1-1 1. Pantalla display: muestra los valores de los parámetros del y la onda pletismográfica 2.
Página 46
Manual de uso para Pulsioxímetro Presionar esta tecla, la pantalla por defecto cambiará para mostrar el pulso (FP) o el índice de la perfusión (IP). Si está en la pantalla del sistema de configuración, presione para mover el cursor hacia arriba o hacia la izquierda y ajustar valores de parámetros.
8. Icono: “SpO ”: SpO Conector del Sensor. 1.2 Producto Nombre y Modelo Nombre: Pulsioxímetro Modelo: PC-66L 1.3 Estructura Está formada por la unidad principal y el sensor de SpO 1.4 Prestaciones Es ligero, de pequeño tamaño y fácil de llevar ...
Manual de uso para Pulsioxímetro 1.5 Uso al que se destina Este oxímetro de pulso está diseñado para verificar y registrar la frecuencia del pulso y la saturación de oxígeno funcional (SpO2) de adultos y niños para uso saludable. Este dispositivo no está...
Figura 2.1. NO abra el compartimiento de la batería a menos que se lo digan por el apoyo técnico de CMI. Figura 2-1 Fijar el soporte, como se muestra en la Figura 2-2.
Página 50
De forma segura ponga la nueva batería en el compartimiento de la batería. NO intentes a menos que se lo digan por el apoyo técnico de CMI. Instrucciones de seguridad para el funcionamiento: No introduzca la batería en el agua, ni en líquidos ni en el ...
Si la máquina no prende correctamente o la batería está dañada, por favor comuniquese con el servicio al cliente de CMI. Para poder prolongar la vida de la batería de litio, por favor preste atención al mantenimiento de la batería.
Página 52
Manual de uso para Pulsioxímetro Instrucciones de Funcionamiento Colocar el dedo dentro del sensor de forma correcta y adecuada. No mueva el dedo y manténgalo en su posición durante toda la lectura. No ponga el dedo mojado directamente dentro del sensor. Evite colocar el sensor en la misma extremidad que tenga colocado el manguito para medir la presión arterial o la perfusión venosa.
Manual de uso para Pulsioximetro 4 Funcionamiento 4.1 Encender/Apagar el Oxímetro Presione la tecla “ ” Energía/Atrás durante 1~2 segundos para encender o apagar el Oxímetro. Durante la medición, si “probe off” (ausencia del sensor) se detecta durante más de un minuto, el dispositivo se apagará...
Página 54
Manual de uso para Pulsioxímetro Descripción: El indicador de sonido " " significa que el equipo está en estado apagado ("mute"), el usuario puede cambiar éste estado pulsando brevemente recall/mute " ". Durante la medición, el evento de superación del límite o sensor desconectado pueden activar la alerta sonora.
Manual de uso para Pulsioxímetro acústica del pulso (el sonido tiende a ser más agudo). Cuanto más bajo es el valor SpO , más baja es la frecuencia del tono de la señal acústica del pulso (el sonido tiende a ser más grave).
Manual de uso para Pulsioxímetro 4.4 Menú Configuración Estando en la pantalla anterior, pulsar durante largo tiempo la tecla " " para entrar en la pantalla del menú de configuración (como se muestra en la Figura 4-3). Figura 4-3 Pantalla Menú de Configuración Descripción de la Pantalla “Setting”: configura los valores de los parámetros, ...
Página 57
Manual de uso para Pulsioxímetro Figura 4-4 Pantalla Sistema de Configuración Instrucciones de Funcionamiento: Patient: Hay dos opciones. "ADU " para adultos, y "PED " para pediátricos. SpO2 Lo-Limit: El rango para ajuste del límite inferior del SpO es de 50%~99%, en escalones del 1%.
Página 58
Manual de uso para Pulsioxímetro PR Hi-Limit: El rango de ajuste para el límite superior de la frecuencia de pulso es de 100~240 ppm. Desde 100 a 150, en escalones de 1 ppm, y desde 150 a 240, en escalones de 5 ppm.
Página 59
Manual de uso para Pulsioxímetro Nota: Las operaciones de configuración de otros parámetros (como HORA; PACIENTE; INTERVALO DE REGISTRO, AHORRO DE ENRGIA, etc) son los mismos que los utilizados para ajustar la fecha. Time: Ajuste de la Hora Recording: El intervalo para registrar los datos (SpO &...
Manual de uso para Pulsioxímetro Volumen: 3 niveles para ajustar: “1”, “2”, y “3” corresponde a los iconos de volumen del altavoz, respectivamente. Los iconos de volumen del altavoz indican el volumen de cualquier sonido generado por la máquina como presionar una tecla, la alarma sobre el límite, el pitido del pulso y una indicación del sondaje oíble.
Manual de uso para Pulsioxímetro 4.4.3 Config. Fábrica En la pantalla de Menú, seleccionar "Config. Fábrica" y a continuación pulsar brevemente la tecla " " para entrar en la pantalla de configuración de fábrica (como se muestra en la Figura 4-6).Presione las teclas de Navegación " "...
Manual de uso para Pulsioxímetro 4.4.5 Versión En la Pantalla menú, seleccionar “VERSION” y a continuación pulsar la tecla “ ” para entrar en la pantalla de la versión (como se muestra en la Figura 4-7). Figura 4-7 4.5 Llamada de Datos En la pantalla de configuración de fábrica, pulsar durante largo tiempo la tecla “...
Página 63
Manual de uso para Pulsioxímetro 4.5.1 Llamada de Datos Elegir el registro en la lista de registros, y pulsar la tecla “ ” (Modo/OK), la pantalla mostrará el gráfico de las tendencias como se muestra en la Figura 4-9A. Intervalo grabación % SpO Gráfica Tendencía...
Manual de uso para Pulsioxímetro Figura 4-9C Pantalla Gráfico Tendencias 4.5.2 Borrado de Datos En la pantalla de menú, pulsar la Tecla “ ” durante largo tiempo y la lista de registros se mostrará. En este momento pulsar de nuevo la tecla “ ”...
Manual de uso para Pulsioximetro 5 Especificaciones Técnicas A. Tipo Pantalla: Color dot-matrix LCD; B. Alimentación: Voltaje alimentación: c.c. 4.5V(Batería de litio recargable), o c.c. 3.6V (batería de Litio recargable) Corriente de funcionamiento: ≤180mA C. Medición SpO Transductor: sensor tipo LED con dos longitudes de onda longitud de onda: Luz roja: 663 nm, luz Infrarroja: 890 nm.
Manual de uso para Pulsioxímetro Temperatura de funcionamiento: 41º~104ºF Humedad de funcionamiento: 30%~80% Presión atmosférica: 70 kPa~106kPa Nota: Los dispositivos portátiles y comunicaciones móviles pueden afectar las prestaciones del oxímetro. G. Actualización Datos El promedio del SpO y la frecuencia de pulso se calcula en base a 8 pulsaciones H.
Manual de uso para Pulsioximetro Tipo BF en las partes aplicadas. Grado de protección contra la entrada de líquidos: IPX1 (protección contra la caída vertical de agua). El equipo no puede utilizarse en entornos ricos en oxígeno o gas inflamable. Modo de uso: uso intensivo.
Página 68
Manual de uso para Pulsioxímetro volumen del altavoz pasa a ser " ". Si esta alerta persiste durante más de 90 segundos, la indicación de superación del límite, y el sonido de alerta se activará de nuevo. Durante la medición, si el sensor se ha salido del dedo o ...
Manual de uso para Pulsioxímetro 7 lista de empaque 1. el oxímetro PC-66L 2. La sonda adulta SpO 3. Un cable de extensión SpO 4. Un estante de metal 5. cable para cargar y adaptador 6. manual para el usuario 7.
Manual de uso para Pulsioxímetro 8 Reparación y Mantenimiento 8.1 Mantenimiento La vida esperada de este producto (no su garantía) es de 5 años. Para poder asegurar un servicio prolongado del equipo, prestar atención al mantenimiento del mismo; Por favor cambie las pilas cuando aparezca el indicador de bajo voltaje.
Manual de uso para Pulsioxímetro 8.2 Instrucciones de Limpieza y Desinfección 5) Limpiar la superficie del sensor con un paño suave impregnado con una solución del 75% de alcohol isopropílico, si se necesita un nivel bajo de desinfección, utilice una solución al 1:10 de lejía. 6) Después de limpiar la superficie con un paño húmedo, dejarlo secar al aire o séquelo con un paño.
Manual de uso para Pulsioxímetro 9 Solución de Problemas Problema Posible Razón Solución Verifica la sonda para ver si la El cable está fuente de la luz está parpadeando máquina desconectado o el una luz roja. Si no, el cable de la seguía circuito está...
Página 73
Manual de uso para Pulsioxímetro El ajuste del favor sensor sensor or el tipo recomendado de acuerdo al tipo de de cuerpo está cuerpo del usuario y el uso. Un incorrecto ajuste impropio del sensor or tipo de cuerpo, puede causar una lectura incorrecta.
¿Puedo usar no ofrece un software para computad software? computadoras Mac. oras PC ¿El software software gratis. gratis? ¿El software Simplemente baja el software viene con la compra del sitio de web de CMI Health: de la máquina?? www.cmihealth.com/download...
Página 75
Manual de uso para Pulsioxímetro ¿Qué sistema CMI Health’s Oximeter Data computadora apoya el Manager software apoya software de la PC? Windows 7/8/10. ¿Cómo puedo CMI Health’s Oximeter Data transferir la data de Manager software ofrece una una computadora a característica para exportar e...
Manual de uso para Pulsioxímetro Apéndice Interpretación de los Símbolos Descripción Símbolos %SpO La saturación de oxígeno Índice de Perfusión Frecuencia de pulso (Unidades: Símbolo pulsaciones por minuto) Bajo voltaje de la batería en la Icono de batería totalmente pantalla cargada Icono de altavoz silenciado Icono volumen altavoz...
Página 77
Manual de uso para Pulsioxímetro Símbolos Descripción Conector sensor SpO Tecla Alimentación/Atrás Tecla Modo/OK Símbolo Tecla Llamar/Silencio / Tecla Navegación en el Interface de Datos panel Marca CE Número de serie Fecha de fabricación Representante Autorizado en la Comunidad Europea Fabricante (incluye dirección) Parte aplicada tipo BF Ver manual del Usuario...
Manual de uso para Pulsioxímetro Conocimientos Generales 1 Significado del SpO El SpO2 es el porcentaje de saturación de oxígeno en la sangre, llamada concentración de O en la sangre; se define por el porcentaje de oxihemoglobina (HbO ) en la hemoglobina total de la sangre arterial.
Página 79
Manual de uso para Pulsioxímetro puede ser determinado. El SpO medido por el oxímetro está en función de la saturación de oxigeno – el porcentaje de la hemoglobina que puede transportar oxígeno. Por contra, los hemoxímetros informan de la fracción de la saturación de oxígeno – un porcentaje de toda la hemoglobina medida, incluyendo la hemoglobina disfuncional tal como la carboxihemoglobina o la metahematoglobina.
Página 80
Manual de uso para Pulsioxímetro 4 Factores que afectan a la precisión del SpO (por interferencias) Colorantes intravasculares como el verde de indocianina o el azul de metileno Exposición a iluminación excesiva, como las lámparas quirúrgicas, lámparas de bilirrubina, luces fluorescentes, lámparas para...
Página 81
Manual de uso para Pulsioxímetro Anormal variación de la oxihemoglobina Enfermedad de Metahemoglobina Sulfohemoglobinemia o existe oclusión arterial próxima al sensor Pulsaciones venosas obvias Pulsaciones periféricas arteriales muy débiles El flujo de sangre periférica es insuficiente ...