Instructions for Use
Please read these instructions carefully before use. Retain this
instruction leaflet for future reference.
Use this product only under the direction of your physician or
qualified medical personnel.
Intended Use
The Multi-Lancet™ Device 2 is a reusable lancing device used to obtain
a capillary blood sample from the fingertip or alternate sites for glucose
monitoring or other tests that require blood. The Multi-Lancet™ Device 2 is
intended for single person use only.
For Your Safety
NOTE
• If the lancing device is being operated by a second person who is providing
assistance to the user, the lancing device should be decontaminated prior
to use by the second person.
• The lancing device is for use by a single person only. It must not be used to
collect blood from more than one person as this poses an infection risk
• Do not share the lancing device with anyone including other family
members! Do not use on multiple patients!
• User should wash hands thoroughly with soap and water after handling the
lancing device.
• All parts of the kit are considered biohazardous and can potentially
transmit infectious diseases, even after you have performed cleaning and
disinfection.
1 "FDA Public Health Notification: Use of Fingerstick Devices on More than One
Person Poses Risk for Transmitting Bloodborne Pathogens: Initial Communication"
(2010)
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm224025.htm
2 "CDC Clinical Reminder: Use of Fingerstick Devices on More than One Person Poses
Risk for Transmitting Bloodborne Pathogens" (2010)
http://www.cdc.gov/injectionsafety/Fingerstick-DevicesBMG.html
Features and Device Configuration
The Multi-Lancet™ Device 2 can draw blood from a finger and a trace
amount of blood from other sites (mainly from the forearm).
Head Cap
Lancet
Fingerstick
Cap (white)
AST Cap
Protective
(clear)
Cap
Head Caps
• There are two head caps; one for fingerstick and one for alternate site
testing (AST).
• Change the cap to match the blood collection method.
• To remove the head cap, push while holding the pinch points.
Adjusting Lancing Depth
The Multi-Lancet™ Device 2 can adjust the lance depth in 7 levels, as
needed. When using the Multi-Lancet™ Device 2 for the first time, select
level 3 for finger samples and level 7 for palm, forearm and upper arm
samples. If you have soft finger skin, start with level 2 and adjust depth
as needed.
• If blood sample is excessive,
turn the dial to a lower setting.
• If blood sample is insufficient,
turn the dial to a higher setting.
CAUTION
Do not turn the dial beyond 1 or 7 as damage to device may occur.
Resetting the Internal Trigger
If the puncture button is inadvertently pressed before
lancing a site, pull the reset lever in the direction of the
arrow until a click is heard. This locks the internal trigger.
Then take your hand off the reset lever.
CAUTION
Do NOT pull the reset lever while touching the puncture button and the
eject lever.
How to Use
Step 1: Remove the head cap
Remove the head cap by pinching and pulling the
cap off the lancing device.
Step 2: Insert the lancet
Insert the lancet by pushing down firmly until lancet
is fully seated.
NOTE
Do not use if protective cap has been removed.
Step 3: Remove the protective cap
Step 4: Attach the head cap
Step 5: Adjust puncture setting
Step 6: Obtain blood sample
NOTE
Consult your blood glucose meter user manual for instructions on
how to obtain a blood sample.
.
1,2
Step 7: Attach the protective cap
Step 8: Dispose of the lancet
Cleaning and Disinfecting Lancing Device
Clean and disinfect the Multi-Lancet™ Device 2 with Clorox
Wipes, EPA Registration number 67619-12. Other brand wipes with EPA
Registration number 67619-12 may also be used.
Reset Lever
Purchase Information for Clorox Healthcare
Clorox
®
websites listed below.
Eject Lever
• Medline........................................................................... 800.633.5463
• DNR Supplies ................................................................ 763.420.7825
• www.amazon.com
• www.walmart.com
If you need assistance or have questions about cleaning and disinfecting,
please contact customer service at 800.566.8558.
NOTE
• Only Clorox
disinfect the lancing device.
• Keep Clorox
Cleaning and Disinfecting Lancing Device
1. Inspect for blood, debris, dust, or lint on the lancing device.
2. To clean the lancing device and head cap, blood/body fluids must be
thoroughly cleaned from the surface of the lancing device and head cap
before disinfecting. Dispense the wipe (Clorox
by gently wiping the outside of the device. Dry thoroughly with a dry cloth.
3. DISINFECT by gently wiping the outside of the device. Allow the device to
remain wet with the disinfectant solution for 1 minute. Dry the device with
a dry cloth.
4. Wash hands with soap and water.
Cleaning and disinfection frequency: ARKRAY has validated a 4-year
use life by exposing the lancing device to 832 wipes (416) cycles with the
disinfectant product listed above. The lancing device should be CLEANED
and DISINFECTED a minimum of once per week and immediately when
blood is visible, for up to 2 cleaning and disinfection cycles per week over
the 4-year use life.
Distributed by
Minneapolis, MN 55439 USA
www.arkrayusa.com
For questions/comments, contact Customer
Service at 800.566.8558, 24 hours a day,
7 days a week, US and Canada.
Made in Philippines
Twist the protective cap off the lancet.
Attach the head cap by pushing it onto the lancing
device.
Adjust the puncture setting for the preferred
puncture depth level (adjustable depths).
After obtaining your blood glucose results, remove
the head cap, then stab the protective cap with the
lancet.
Eject the used lancet using the eject lever and
dispose of it in a safe manner to avoid accidental
lancet stick injuries.
wipes can be ordered from these national distributors and on the
wipes or EPA 67619-12 should be used to clean and
®
wipes out of reach of children.
®
Germicidal
®
Bleach Germicidal Wipes
®
) from its package. CLEAN
®
P/N 5174-01
Rev 03/15