Filter Specifications
Standard Filter ............................................................................................................................................................................................................................................ > 3.0 micron particles
Optional Fine Particle Filter ........................................................................................................................................................................................................................ > 0.3 micron particles
Dynamic Short-Term Pressure Accuracy with and without DV6HH Humidifier (Per ISO 17510-1:2007)
Waveform
Volume [mL]
Sinusoidal cycle with
500
I:E = 1:1
Maximum Flow-Rate (per ISO 17510-1:2007)
Measured pressure at the patient connection port (cmH
Average flow at the patient connection port (l/min)
Long Term Static Pressure Accuracy .....................................................................................................................................................................................................................+/- 0.5 cmH
Additional Specifications
Equipment classification with respect to protection from electric shock ........................................................................................................................................................................... Class II
Degree of protection from electric shock .................................................................................................................................................................................................... Type BF Applied Part
Degree of protection against ingress of liquids ................................................................................................. IP21 Ingress Protection – Protected against finger access to hazardous parts;
Mode of operation ....................................................................................................................................................................................................................................................... Continuous
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide. See instructions for use with supplemental oxygen.
US Patents
SmartCode Technology is protected under ................................................................................................................................................................................................... US Patent 8649510
Approvals
Meets RTCA/DO-160 .................................................................................................................................................... DO-160 Section 21 Category M for battery operation only Airline Use
SE-DV64-1
Set Pressure
Breath Rate [min-1]
10, 15, 20
3.0 cmH
O
2
O)
2.9
2
78.2
Pressure Accuracy
[cmH
O]
2
3.0
7.0
12.0
16.0
20.0
Test Pressures
7.0 cmH
O
12.0 cmH
2
6.6
123.5
CPAP only
CPAP with Humidifier
[Pk-Pk cmH
O]
2
0.5
0.5
1.0
1.0
1.0
O
16.0 cmH
O
2
2
11.4
15.0
166.8
157.7
protected against vertically falling water drops
Pressure Accuracy
[Pk-Pk cmH
O]
2
0.5
0.5
1.0
1.0
1.0
20.0 cmH
O
2
19.0
138.0
O
2
EN - 29