HEINE OMEGA 600 Manual Del Usuario página 48

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 15
Test
Electrostatic Discharge
(IEC 61000-4-2)
Radiated RF EM fields
(IEC 61000-4-3)
Electrical fast transients / bursts
(IEC 61000-4-4)*
Surges
(IEC 61000-4-5)*
Conducted disturbances
induced by RF fields
(IEC 61000-4-6)*
Proximity fields from RF wireless
communications equipment
(IEC 61000-4-3)
Power frequency magnetic fields
(IEC 61000-4-8)
Voltage dips
(IEC 61000-4-11)*
Short interruptions
(IEC 61000-4-11)*
*n/a: "Not applicable" in the internally powered mode
48/52
Immunity test levels
IEC 60601-1-2 test levels
Contact Discharge: ±8 kV
Air Discharge: ±2 kV, ±4 kV, ±8 kV, ±15 kV
3 V/m
80 – 2700 MHz
80 % AM at 1kHz
± 2 kV
100 kHz repetition frequency
± 0.5 kV, ± 1 kV
3 V
0.15 MHz – 80 MHz
6 V in ISM bands between 0.15 MHz and 80 MHz
80 % AM at 1 kHz
385 MHz; Pulse Modulation: 18 Hz; 27 V/m
450 MHz, FM: ±5 Hz deviation: 1 kHz sine; 28 V/m
710, 745, 780 MHz; Pulse Modulation: 217 Hz; 9 V/m
810, 870, 930 MHz; Pulse Modulation: 18 Hz; 28 V/m
1720, 1845, 1970 MHz; Pulse Modulation: 217 Hz; 28 V/m
2450 MHz; Pulse Modulation: 217 Hz; 28 V/m;
5240, 5500, 5785 MHz; Pulse Modulation: 217 Hz; 9 V/m
30 A/m; 50Hz or 60 Hz
0 % U
; 0.5 cycle; at 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
T
0 % U
; 1 cycle and 70 % U
T
0 % U
; 250/300 cycles
T
MED 234766 2021-03-26
Test level
Compliance test levels
30 A/m; 60 Hz
; 25/30 cycles Single phase: at 0°
T