▒ Change the type of blood circulation display
< V Type >
• When you compress the chest of the manikin over 5㎝ deep, for longer than 3 seconds,
the blood circulation display will change from <V type> to <O type> (see illustration
above). If the manikin is switched o , it will always default back to <v type> when it is
turned back on.
▒ Note
• If the chest compression is not detected for 2 seconds, the blood circulation indication lights, the compression depth indication lights and the normal circulation indication light will
be turned o and the manikin will enter stand-by mode. When in stand-by mode, the power status indicator lamp will ash green.
• When the user carries out chest compressions again while the manikin is in stand-by mode, the chest compressions will be detected automatically and the indicators will operate
again. However, if left in stand-by mode for 30 minutes, the manikin will switch itself o to conserve the batteries. To switch the manikin on again, slide the power switch to the 'o '
position for more than 2 seconds and then switch it on again.
• While in stand-by mode, the power consumption is very low. However, we recommend that you switch o the manikin when not in use.
• If the battery is low, the power status indicator lamp will ash red. At this time, replace the battery with a new one.
< O Type >
▒ Change the clicker setting
< ON >
• Turn the 'clicker' on and o using the switch as shown. The clicker makes a "clicking"
sound when the chest is compressed more than 5㎝.
< OFF >
5